TLD, diode and monte carlo dosimetry of an Ir-192 source for high dose-rate brachytherapy

被引:83
作者
Kirov, AS
Williamson, JF
Meigooni, AS
Zhu, Y
机构
[1] Mallinckrodt Inst. of Radiol., Washington Univ. Sch. of Med., St. Louis, MO
关键词
D O I
10.1088/0031-9155/40/12/002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Very few dosimetry data are available for the current generation of high-dose-rate (HDR) Ir-192 sources, which have broad application in remotely afterloaded brachytherapy. We have measured the two-dimensional dose rate distribution around a microSelectron-HDR source and used the results to validate Monte Carlo simulations. Thermoluminescent dosimeters (TLDs) in solid-water phantoms were used to measure the transverse-axis dose rates in the distance range 0.5-10 cm and the polar dose-rate profiles at 1.5, 3 and 5 cm distance from the source. At close distances, 2-40 mm from the HDR source, we performed transverse axis dose-rate measurements with a Si diode in water. We performed diode measurements at the same distances also for a pulsed dose-rate (PDR) source to compare the results for Ir-192 sources With different encapsulation. Both the HDR and the PDR sources were decayed, separated from their cables and calibrated prior to the measurements. The measured dose rates were compared with Monte Carlo photon transport calculations, which realistically modelled the experimental and source geometry at each measurement point. Agreement between Monte Carlo photon transport absolute dose-rate calculations and measurements was, on average, within 5%. From the transverse-axis experimental data, we deduced a value for the dose-rate constant Lambda(0) of Ir-192 HDR sources of 1.14 cGy h(-1) U-1 +/- 5%. This value agrees within the experimental error with the Monte Carlo estimate of 1.115 cGy h(-1) U-1 +/- 0.5%. Excellent agreement with previously measured anisotropy functions was observed. Higher anisotropy is observed for the point at 0 degrees along the source cable for which no previous data have been reported.
引用
收藏
页码:2015 / 2036
页数:22
相关论文
共 36 条
[1]   AUTORADIOGRAPHY FOR I-125 SEEDS [J].
ALBERTI, W ;
DIVOUX, S ;
POTHMANN, B ;
TABOR, P ;
HERMANN, KP ;
HARDER, D .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (05) :881-884
[2]  
[Anonymous], 1983, MED PHYS, V10, P741
[3]  
Attix F.H., 1986, INTRO RADIOLOGICAL P, P607, DOI 10.1002/9783527617135
[4]  
BALTAS D, 1993, 3 ACT SPEC REP
[5]  
Cameron JR, 1968, THERMOLUMINESCENT DO
[6]  
CERRA T, 1990, ENDOCURIE HYPER ONCO, V6, P71
[7]  
CHIUTSAO ST, 1987, MED PHYS, V14, P478
[8]  
DAS RK, 1995, COMMUNICATION
[9]  
GLASGOW G P, 1979, Medical Physics (Woodbury), V6, P49, DOI 10.1118/1.594551
[10]   CALIBRATION OF IR-192 HIGH-DOSE-RATE AFTERLOADING SYSTEMS [J].
GOETSCH, SJ ;
ATTIX, FH ;
PEARSON, DW ;
THOMADSEN, BR .
MEDICAL PHYSICS, 1991, 18 (03) :462-467