ONLINE PRECONCENTRATION AND FLOW-ANALYSIS FOURIER-TRANSFORM INFRARED DETERMINATION OF CARBARYL

被引:24
作者
GARRIGUES, S [1 ]
VIDAL, MT [1 ]
GALLIGNANI, M [1 ]
DELAGUARDIA, M [1 ]
机构
[1] UNIV VALENCIA,DEPT ANALYT CHEM,E-46100 BURJASSOT,SPAIN
关键词
CARBARYL WATER ANALYSIS; FOURIER TRANSFORM INFRARED; SOLID PHASE EXTRACTION; ONLINE ANALYSIS;
D O I
10.1039/an9941900659
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new strategy to improve the limits of detection in Fourier transform Infrared (FTIR) has been developed. The proposed method consists of preconcentration of samples, using conventional solid phase extraction cartridges, and online elution with dichloromethane. Absorbance measurements are obtained by FTIR. Carbaryl has been used as a test molecule with on-line measurements carried out at 1746 cm(-1). This procedure permits (i) field sampling, (ii) stabilization of pesticides, and (iii) fast determination of carbaryl in the laboratory. The sensitivity obtained with this procedure corresponds to 0.0444 area units mu g(-1) cm(-1), and the limit of detection, to 15 mu g of carbaryl, which, for a sampling volume of 1 l, corresponds to 15 mu g l(-1) of this pesticide. The reproducibility of three independent analyses of a well water sample containing 195 mu g l(-1) of carbaryl corresponds to a relative standard deviation of 8.6%.
引用
收藏
页码:659 / 664
页数:6
相关论文
共 31 条
[31]   THE USE OF SOLID-PHASE EXTRACTION COLUMNS TO EFFECT SIMPLE OFFLINE LC/MS, LC/NMR, AND LC/FTIR [J].
WILCOX, CD ;
PHELAN, RM .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1986, 24 (04) :130-133