MAGNETOSTIMULATION IN MRI

被引:72
作者
IRNICH, W [1 ]
SCHMITT, F [1 ]
机构
[1] SIEMENS AG,DEPT MR DEV,ERLANGEN,GERMANY
关键词
MRI; MAGNETIC FIELD; ELECTROSTIMULATION; MAGNETOSTIMULATION;
D O I
10.1002/mrm.1910330506
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In national and international bodies, there is active discussion of appropriate safety regulations of levels of magnetic field strength in MRI. Present limits are usually expressed in terms of the switching rate dB/dt, but the validity of this is open to debate. Application of the fundamental law of electrostimulation is well-established, both on theoretical and experimental grounds. Application of this law, in combination with Maxwell's law, yields a very simple equation that we call the fundamental law of magnetostimulation. This law has the hyperbolic form of a strength-duration curve and allows an estimation of the lowest possible value of the magnetic flux density capable of stimulating nerves and muscles. Calculations prove that the threshold for heart excitation is much higher than those for nerve and muscle stimulations. Experimental results from us and other authors confirm the correctness of the derived laws for magnetostimulation. In light of these findings, proposed safety limits should be reconsidered.
引用
收藏
页码:619 / 623
页数:5
相关论文
共 11 条
[1]   PHYSIOLOGICAL-EFFECTS OF FAST OSCILLATING MAGNETIC-FIELD GRADIENTS [J].
BUDINGER, TF ;
FISCHER, H ;
HENTSCHEL, D ;
REINFELDER, HE ;
SCHMITT, F .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1991, 15 (06) :909-914
[2]   SENSORY STIMULATION BY TIME-VARYING MAGNETIC-FIELDS [J].
COHEN, MS ;
WEISSKOFF, RM ;
RZEDZIAN, RR ;
KANTOR, HL .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (02) :409-414
[3]   OPTIMAL TRUNCATION OF DEFIBRILLATION PULSES [J].
IRNICH, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (04) :673-688
[4]   THE FUNDAMENTAL LAW OF ELECTROSTIMULATION AND ITS APPLICATION TO DEFIBRILLATION [J].
IRNICH, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (11) :1433-1447
[5]  
Irnich W., 1994, MAGMA, V2, P43, DOI 10.1007/BF01709799
[6]  
KONING G, 1975, MED BIOL ENG COMPUT, V313, P388
[7]  
LAPICQUE L, 1989, SOC BIOL, V77, P280
[8]   LIMITS TO NEURAL STIMULATION IN ECHO-PLANAR IMAGING [J].
MANSFIELD, P ;
HARVEY, PR .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (06) :746-758
[9]  
REILLY JP, 1992, ELECTRICAL STIMULATI
[10]  
SILNY J, 1987, G HEINEMANN UFER, V130