IN-VIVO VALIDATION OF MAGNETIC-RESONANCE BLOOD-VOLUME FLOW MEASUREMENTS WITH LIMITED SPATIAL-RESOLUTION IN SMALL VESSELS

被引:114
作者
HOFMAN, MBM
VISSER, FC
VANROSSUM, AC
VINK, GQM
SPRENGER, M
WESTERHOF, N
机构
[1] FREE UNIV AMSTERDAM, DEPT PHYSIOL, AMSTERDAM, NETHERLANDS
[2] FREE UNIV AMSTERDAM, DEPT CLIN PHYS & ENGN, AMSTERDAM, NETHERLANDS
[3] FREE UNIV AMSTERDAM, DEPT CLIN ANIM EXPET, AMSTERDAM, NETHERLANDS
[4] FREE UNIV AMSTERDAM, DEPT CARDIOL, AMSTERDAM, NETHERLANDS
关键词
VOLUME FLOW; MRI; ACCURACY; FLOW QUANTIFICATION;
D O I
10.1002/mrm.1910330606
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The accuracy of magnetic resonance phase contrast volume flow measurements in small blood vessels is expected to be smaller than in large vessels, because of partial volume effects at the vessel boundary, Accuracy was validated in the dog femoral artery, diameter 3.5 +/- 0.7 mm, using an ultrasonic transit-time flowmeter (TT). The number of pixels per vessel diameter (N-D) ranged from 1,6 to 4.8, The vessel cross-section was determined using a threshold in the magnitude image. Between the two methods the correlation coefficient was 0.95 (range 10-200 ml/min), The proportional difference (PD), (Q(TT) - Q(MR))/1/2(Q(TT) + Q(MR)), was 0.8%, showing no systematic difference between the methods, The PDs standard deviation was 27%, and 19% for flow rates above 30 ml/min, Only a significant decrease of the PDs variance was found at the highest N-D values, suggesting other sources of error than partial volume effects, It is concluded that with an N-D value of about 3, accurate blood Volume flow rates can be determined.
引用
收藏
页码:778 / 784
页数:7
相关论文
共 32 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   SIGNAL-TO-NOISE IN PHASE-ANGLE RECONSTRUCTION - DYNAMIC-RANGE EXTENSION USING PHASE REFERENCE OFFSETS [J].
CONTURO, TE ;
SMITH, GD .
MAGNETIC RESONANCE IN MEDICINE, 1990, 15 (03) :420-437
[3]   ACCURACY OF COLOR DOPPLER ULTRASOUND VELOCITY-MEASUREMENTS IN SMALL VESSELS [J].
DEANE, CR ;
FORSBERG, F ;
THOMAS, N ;
ROBERTS, VC .
JOURNAL OF BIOMEDICAL ENGINEERING, 1991, 13 (03) :249-254
[4]   RENAL-ARTERY BLOOD-FLOW - QUANTITATION WITH PHASE-CONTRAST MR-IMAGING WITH AND WITHOUT BREATH-HOLDING [J].
DEBATIN, JF ;
TING, RH ;
WEGMULLER, H ;
SOMMER, FG ;
FREDRICKSON, JO ;
BROSNAN, TJ ;
BOWMAN, BS ;
MYERS, BD ;
HERFKENS, RJ ;
PELC, NJ .
RADIOLOGY, 1994, 190 (02) :371-378
[5]  
DROST CJ, 1978, 31ST P ANN C ENG MED, P183
[6]   INVIVO VALIDATION OF MR VELOCITY IMAGING [J].
FIRMIN, DN ;
NAYLER, GL ;
KLIPSTEIN, RH ;
UNDERWOOD, SR ;
REES, RSO ;
LONGMORE, DB .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1987, 11 (05) :751-756
[7]   A PULSED ULTRASONIC FLOWMETER [J].
FRANKLIN, DL ;
BAKER, DW ;
ELLIS, RM ;
RUSHMER, RF .
IRE TRANSACTIONS ON MEDICAL ELECTRONICS, 1959, 6 (04) :204-206
[8]  
GROEN JP, 1988, 7TH P ANN M SOC MAGN, P906
[9]   NMR SPIN-ECHO FLOW MEASUREMENTS [J].
GROVER, T ;
SINGER, JR .
JOURNAL OF APPLIED PHYSICS, 1971, 42 (03) :938-&
[10]   MR VASCULAR IMAGING WITH A FAST GRADIENT REFOCUSING PULSE SEQUENCE AND REFORMATTED IMAGES FROM TRANSAXIAL SECTIONS [J].
GULLBERG, GT ;
WEHRLI, FW ;
SHIMAKAWA, A ;
SIMONS, MA .
RADIOLOGY, 1987, 165 (01) :241-246