AN AUTOMATED RESEARCH FACILITY FOR MEASURING THERMOLUMINESCENCE EMISSION-SPECTRA USING AN OPTICAL MULTICHANNEL ANALYZER

被引:29
作者
PITERS, TM
MEULEMANS, WH
BOS, AJJ
机构
[1] Interfaculty Reactor Institute, Delft University of Technology, NL-2629 JB Delft
关键词
D O I
10.1063/1.1144423
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A facility for research into the mechanism of thermoluminescence (TL) is described. The facility comprises three units: an annealing oven, an irradiator, and a TL-emission spectrometer. Crystals or hot-pressed chips can be moved from and to the mentioned units by an automated sample changer. All units operate automatically and are controlled by a personal computer program. The spectrometer is based on a dispersive grating and an intensified diode array (512 active elements) and covers the 200-800-nm wavelength range. The wavelength resolution of the spectrometer is 6 nm when a 25-mum-wide entrance slit is used and 29 nm when a 1-mm-wide entrance slit is used. The sensitive spectrometer could measure emission spectra of CaSO4:Dy (TLD-200) irradiated at an absorbed dose as low as 3 mGy at a signal-to-noise ratio of 10:1 for LiF:Mg,Ti (TLD-100) this ratio was obtained at 75 mGy. A detailed description is given how measured data can be related to spectra predicted by a model, taking into account all system aberrations. Spectra of LiF:Mg,Ti (TLD-100) irradiated to an absorbed dose of 5 Gy are analyzed according to the Franck-Condon model for light emission. Two emission bands with peak energies of 3.01 and 2.60 eV at 463 K have been found.
引用
收藏
页码:109 / 117
页数:9
相关论文
共 18 条
[11]  
KANEMAKI M, 1991, NUCL TRACKS RAD MEAS, V18, P81
[12]   THE FRANCK-CONDON PRINCIPLE AND ITS APPLICATION TO CRYSTALS [J].
LAX, M .
JOURNAL OF CHEMICAL PHYSICS, 1952, 20 (11) :1752-1760
[13]  
MARTIN BR, 1971, STATISTICS PHYSICIST, P85
[14]  
Mckeever SWS, 2011, THERMOLUMINESCENCE S, DOI [10.1017/CBO9780511564994, DOI 10.1017/CBO9780511564994]
[15]   THERMO-LUMINESCENCE EMISSION SPECTROMETER [J].
PRESCOTT, JR ;
FOX, PJ ;
AKBER, RA ;
JENSEN, HE .
APPLIED OPTICS, 1988, 27 (16) :3496-3502
[16]  
SALMON L, 1972, AERER7129 REP
[17]  
TOWNSEND PD, 1983, RADIAT EFF DEFECT S, V72, P245, DOI 10.1080/00337578308218649
[18]   SPECTRAL MEASUREMENT DURING THERMOLUMINESCENCE - AN ESSENTIAL REQUIREMENT [J].
TOWNSEND, PD ;
KIRSH, Y .
CONTEMPORARY PHYSICS, 1989, 30 (05) :337-354