SCALING LIMITATIONS OF SILICON MULTICHANNEL RECORDING PROBES

被引:112
作者
NAJAFI, K
JI, J
WISE, KD
机构
[1] Center for Integrated Sensors and Circuits, Solid-State Electronics Laboratory, Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor
关键词
D O I
10.1109/10.43605
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes the scaling limitations of multichannel recording probes fabricated for use in neurophysiology using silicon integrated circuit technologies. Scaled silicon probe substrates 8 μm thick and 16 μm wide can be fabricated using boron etch-stop techniques. Theoretical expressions for calculating the thickness and width of silicon substrates have been derived and agree closely with experimental results. The effects of scaling probe dimensions on its strength and stiffness are described. The probe shank dimensions can be designed to vary the strength and stiffness for different applications. The scaled silicon substrates have a fracture stress of about 2 x 10 10 dyn/cm2, which is about six times that of bulk silicon, and are strong and very flexible. Scaling the feature sizes of recording electrode arrays down to 1 μm is possible with less than 1 percent electrical crosstalk between channels. © 1990 IEEE
引用
收藏
页码:1 / 11
页数:11
相关论文
共 13 条