EVALUATION OF A SMALL SODA LIME CANISTER IN A TO-AND-FRO SYSTEM

被引:5
作者
KADIM, MY
LOCKWOOD, GG
CHAKRABARTI, MK
WHITWAM, JG
机构
[1] Department of Anaesthetics, Hammersmith Hospital, London, W12 0NN, DuCane Road
关键词
CARBON DIOXIDE; ABSORPTION; EQUIPMENT; ABSORBERS;
D O I
10.1111/j.1365-2044.1991.tb09857.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The main disadvantages of the to‐and‐fro system (the bulky canister and the progressive increase in apparatus deadspace) may be overcome by the use of a smaller canister. In this laboratory study, we have evaluated a 160 g canister in a low‐flow to‐and‐fro system (fresh gas flow 1 litre I minute). Two carbon dioxide productions of 150 and 200 ml/minute were simulated. The mean times to exhaustion, defined here as a 0.5 kPa rise in end‐tidal Pco2, were 112 and 79 minutes in the 150 and 200 ml/minute carbon dioxide groups respectively. Ventilation to normacapnia or hypocapnia did not affect the times to exhaustion. The soda lime absorbed 16 litres of carbon dioxide before exhaustion, and this was not affected by minute volume or carbon dioxide production. A small soda lime canister is suitable for carbon dioxide absorption in a low‐flow to‐and‐fro system for ventilated adults. Copyright © 1991, Wiley Blackwell. All rights reserved
引用
收藏
页码:952 / 956
页数:5
相关论文
共 16 条
[1]  
ADRIANI J, 1946, CHEM ANAESTHESIA, P92
[2]  
Adriani John, 1941, ANES THESIOLOGY, V2, P1, DOI 10.1097/00000542-194101000-00001
[3]   THE VENTURI ANESTHESIA CIRCUIT .3. CARBON-DIOXIDE ELIMINATION [J].
ANDERSEN, APD ;
CHRISTENSEN, KJS ;
JORGENSEN, S .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1985, 29 (03) :276-279
[4]   A NEW VALVELESS ALL-PURPOSE VENTILATOR - DESCRIPTION AND LABORATORY EVALUATION [J].
CHAKRABARTI, MK ;
WHITWAM, JG .
BRITISH JOURNAL OF ANAESTHESIA, 1983, 55 (10) :1005-1015
[5]   EVALUATION OF LUNG VENTILATOR PERFORMANCE ANALYZER [J].
CHAKRABARTI, MK ;
SYKES, MK .
ANAESTHESIA, 1976, 31 (04) :521-528
[6]   THE VENTURI ANESTHESIA CIRCUIT .2. CARBON-DIOXIDE PRODUCTION AND GAS-FLOW REQUIREMENTS [J].
CHRISTENSEN, KJS ;
ANDERSEN, APD ;
JORGENSEN, S .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1985, 29 (03) :273-275
[7]   EFFECTS OF INFLOW OVERFLOW AND VALVE PLACEMENT ON ECONOMY OF CIRCLE SYSTEM [J].
EGER, EI ;
ETHANS, CT .
ANESTHESIOLOGY, 1968, 29 (01) :93-&
[8]   600 GRAM CO2 ABSORPTION CANISTER - EXPERIMENTAL-STUDY [J].
JORGENSEN, B ;
JORGENSEN, S .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1977, 21 (05) :437-444
[9]   THE VENTURI ANESTHESIA CIRCUIT .1. AN ALL-PURPOSE BREATHING SYSTEM FOR ANESTHESIA [J].
JORGENSEN, S ;
HANSEN, LK .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1985, 29 (03) :269-272
[10]   A LOW-FLOW TO-AND-FRO SYSTEM - LABORATORY STUDY OF MIXING OF ANESTHETIC AND DRIVING GASES DURING MECHANICAL VENTILATION [J].
KADIM, MY ;
LOCKWOOD, GG ;
CHAKRABARTI, MK ;
WHITWAM, JG .
ANAESTHESIA, 1991, 46 (11) :948-951