ION DETECTION LIMITATIONS TO MASS RESOLUTION IN MATRIX-ASSISTED LASER DESORPTION TIME-OF-FLIGHT MASS-SPECTROMETRY

被引:4
作者
BROWN, RS
GILFRICH, NL
机构
[1] Department of Chemistry, Colorado State University, Fort Collins, Colorado
关键词
D O I
10.1002/rcm.1290061112
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The ion detection process in a discrete-dynode electron multiplier can result in significant mass resolution losses in time-of-flight mass spectrometry (TOF-MS) for higher mass-to-charge (m/z) ion species. This resolution loss is attributed to propagation time delays and signal broadening in the ion detector. This is presumed to be due to the generation of a distribution of secondary ion species produced initially upon impact of a primary ion with the first dynode surface of the ion detector. Comparisons are made between the signals produced by a standard discrete dynode ion detector (which amplifies the negatively charged species produced by impact of a primary ion) and a detector modified to respond to only the positively charged secondary ion species produced by a primary ion impact. Ion signals for higher m/z ions with the standard detector geometry are seen to be due to a narrow signal component, most likely due to the generation of secondary electrons and/or very low mass secondary ions (H-), and a broad signal component, apparently due to secondary ions which take significant amounts of time to traverse the low potential fields between the first and second detector dynode. This results in ion signal tailing for higher m/z ion species. Numerical subtraction of the ion signal obtained with the standard and modified detector geometries (singly protonated molecular ion species of equine myoglobin) results in an improvement in mass resolution, such that a new adduct ion species (from trifluoroacetic acid) can be resolved.
引用
收藏
页码:690 / 696
页数:7
相关论文
共 21 条
[1]  
Beavis R C, 1989, Rapid Commun Mass Spectrom, V3, P436, DOI 10.1002/rcm.1290031208
[2]  
Beavis R C, 1989, Rapid Commun Mass Spectrom, V3, P233, DOI 10.1002/rcm.1290030708
[3]   VELOCITY DISTRIBUTIONS OF INTACT HIGH MASS POLYPEPTIDE MOLECULE IONS PRODUCED BY MATRIX ASSISTED LASER DESORPTION [J].
BEAVIS, RC ;
CHAIT, BT .
CHEMICAL PHYSICS LETTERS, 1991, 181 (05) :479-484
[4]   HIGH-ACCURACY MOLECULAR MASS DETERMINATION OF PROTEINS USING MATRIX-ASSISTED LASER DESORPTION MASS-SPECTROMETRY [J].
BEAVIS, RC ;
CHAIT, BT .
ANALYTICAL CHEMISTRY, 1990, 62 (17) :1836-1840
[5]  
BORWN RS, UNPUB
[6]  
BROWN R, UNPUB
[7]   DESIGN AND PERFORMANCE OF A MATRIX-ASSISTED LASER DESORPTION TIME-OF-FLIGHT MASS-SPECTROMETER UTILIZING A PULSED NITROGEN LASER [J].
BROWN, RS ;
GILFRICH, NL .
ANALYTICA CHIMICA ACTA, 1991, 248 (02) :541-552
[8]   OPTIMIZING SIGNAL AND MASS RESOLUTION FOR MATRIX-ASSISTED LASER DESORPTION UTILIZING A LINEAR TIME-OF-FLIGHT MASS-SPECTROMETER [J].
BROWN, RS ;
GILFRICH, NL .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1992, 6 (11) :697-701
[9]  
BROWN RS, UNPUB APPL SPECTROSC
[10]   SECONDARY-ELECTRON EMISSION INDUCED BY IMPACT OF LOW-VELOCITY MOLECULAR-IONS ON A MICROCHANNEL PLATE [J].
GENO, PW ;
MACFARLANE, RD .