AUTOMATED BLOOD-PRESSURE MEASUREMENTS IN LABORING WOMEN - ARE THEY RELIABLE

被引:16
作者
MARX, GF [1 ]
SCHWALBE, SS [1 ]
CHO, E [1 ]
WHITTY, JE [1 ]
机构
[1] YESHIVA UNIV ALBERT EINSTEIN COLL MED,DEPT OBSTET & GYNECOL,BRONX,NY 10461
关键词
BLOOD PRESSURE IN PARTURIENTS; AUSCULTATORY VS OSCILLATORY METHOD; PREGNANCY-INDUCED HYPERTENSION; OBSTETRIC ANESTHESIA;
D O I
10.1016/S0002-9378(12)90822-4
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
OBJECTIVE: In laboring women a consistent difference has become evident between measurements obtained with an automated blood pressure device and those obtained with the auscultatory method. A prospective study was designed to assess the concordance of these two methods. STUDY DESIGN: Three sets of brachial blood pressure measurements were made by both oscillatory and auscultatory techniques in 30 women in labor, 20 term pregnant women not in labor, and 20 nonpregnant volunteers. RESULTS: In the nonlaboring women and the nonpregnant controls there was satisfactory agreement between the results of the two methods of measurement. In the parturients systolic pressures were consistently and significantly higher and diastolic pressures consistently and significantly lower with the oscillatory compared with the auscultatory method, but mean arterial pressures were not different. CONCLUSION: In laboring women there is a discrepancy between systolic and diastolic pressures obtained by the auscultatory versus the oscillatory method of measurement, although mean pressures are not significantly different. We suggest that during labor the diagnoses of hypertension and hypotension be based on the mean rather than the systolic or diastolic pressure.
引用
收藏
页码:796 / 798
页数:3
相关论文
共 15 条
[1]  
BORROW KM, 1982, AM HEART J, V103, P879
[2]   RELIABILITY OF THE DINAMAP NONINVASIVE MONITOR IN THE MEASUREMENT OF BLOOD-PRESSURE OF III ASIAN NEWBORNS [J].
CHIA, F ;
ANG, AT ;
WONG, TW ;
TAN, KW ;
FUNG, KP ;
LEE, J ;
KHIN, K .
CLINICAL PEDIATRICS, 1990, 29 (05) :262-267
[3]  
FRIESEN RH, 1981, ANESTH ANALG, V60, P742
[4]   PRESSURE-FLOW RELATIONSHIP IN NONPREGNANT UTERINE VASCULAR BED [J].
GREISS, FC ;
ANDERSON, SG .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1974, 118 (06) :763-772
[5]  
Korotkoff N., 1905, IZV VENNO MED AKAD, P365, DOI DOI 10.1097/00004872-200501000-00001
[6]  
KROTKOFF NS, 1941, B NY ACAD MED, V17, P877
[7]  
MARX GF, 1991, ANESTH ANALG, V72, P562
[8]   IMPACT OF MEAN ARTERIAL-PRESSURE IN MIDDLE TRIMESTER UPON OUTCOME OF PREGNANCY [J].
PAGE, EW ;
CHRISTIANSON, R .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1976, 125 (06) :740-746
[9]   NON-INVASIVE AUTOMATIC-DETERMINATION OF MEAN ARTERIAL-PRESSURE [J].
RAMSEY, M .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1979, 17 (01) :11-18
[10]  
RIVAROCCI S, 1886, GAZ MED TORNIO, V47, P981