RADIATION-DOSE REDUCTION IN PEDIATRIC FLUOROSCOPY USING ADDED FILTRATION

被引:23
作者
NICHOLSON, RA
THORNTON, A
AKPAN, M
机构
[1] Department of Radiology, St Mary's Hospital, London W2 1NY, Praed Street
关键词
D O I
10.1259/0007-1285-68-807-296
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Studies on a water phantom indicate that the insertion of extra tube filtration combined with the removal of the antiscatter grid can reduce the total energy imparted from paediatric fluoroscopy by a factor of more than four when automatic exposure control is used. The omission of the grid is responsible for up to 40% of the reduction while the addition of 0.7 mm steel filtration is found to account for a further reduction in energy imparted of between 57% and 70%, for depths of water simulating small patients of 6-16 cm thickness. For all contrast paediatric examinations the reduction in image quality was considered to be small and the diagnostic information unimpaired with the filter in place.
引用
收藏
页码:296 / 300
页数:5
相关论文
共 10 条
  • [1] RESULTS OF A SURVEY OF DOSES TO PEDIATRIC-PATIENTS UNDERGOING COMMON RADIOLOGICAL EXAMINATIONS
    CHAPPLE, CL
    FAULKNER, K
    LEE, REJ
    HUNTER, EW
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (771) : 225 - 231
  • [2] RADIATION-DOSES TO PEDIATRIC-PATIENTS UNDERGOING LESS COMMON RADIOLOGICAL PROCEDURES INVOLVING FLUOROSCOPY
    CHAPPLE, CL
    FAULKNER, K
    LEE, REJ
    HUNTER, EW
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (789) : 823 - 827
  • [3] Cordt I, 1990, Schweiz Monatsschr Zahnmed, V100, P1160
  • [4] DOSE REDUCTION BY THE USE OF ERBIUM FILTRATION IN A GENERAL RADIOGRAPHIC ROOM
    CRANAGE, RW
    HOWARD, CJ
    WELSH, AD
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (771) : 232 - 237
  • [5] FLUOROSCOPY WITHOUT THE GRID - METHOD OF REDUCING THE RADIATION-DOSE
    DRURY, P
    ROBINSON, A
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1980, 53 (626) : 93 - 99
  • [6] HEINRICH H, 1976, ANN RADIOL, V19, P57
  • [7] Kapa S F, 1990, Dentomaxillofac Radiol, V19, P67
  • [8] MACDONALDJANKOW.DS, 1991, BRIT J RADIOL, V64, P524
  • [9] 1992, NATIONAL PROTOCOL PA
  • [10] 1988, GUIDANCE NOTES PROTE