COMPARISON OF SINGLE-PHOTON AND DUAL-ENERGY X-RAY ABSORPTIOMETRY OF THE RADIUS

被引:23
作者
NELSON, D [1 ]
FEINGOLD, M [1 ]
MASCHA, E [1 ]
KLEEREKOPER, M [1 ]
机构
[1] HENRY FORD HOSP, DIV BIOSTAT RES EPIDEMIOL & COMP, DETROIT, MI 48202 USA
来源
BONE AND MINERAL | 1992年 / 18卷 / 01期
关键词
DUAL-ENERGY X-RAY ABSORPTIOMETRY; SINGLE-PHOTON ABSORPTIOMETRY; BONE MINERAL DENSITY; RADIUS;
D O I
10.1016/0169-6009(92)90801-J
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We compared dual-energy x-ray absorptiometry (DXA) for measurement of the radius with the conventional single-photon absorptiometry (SPA) method. To evaluate reproducibility, 34 healthy male and female subjects were measured twice eacy by both methods. While the instruments measured bone mineral content (BMC) similarly, projected area was consistently lower by SPA and therefore bone mineral density (BMD) was higher by an average of 10%. We report similar coefficients of variation for the two methods, which are 0.8% (SPA) and 0.7% (DXA) for BMC (P = 0.71) and 0.8% (SPA) and 1.4% (DXA) for BMD (P = 0.02). We also evaluated the relationship between the methods with data from 196 clinic patients who were measured once each on both instruments. The BMD measurements from the two instruments were highly correlated (r = 0.97) in these patients, which permits the conversion of databases from SPA to DXA-equivalents. We conclude that DXA can replace SPA for radial bone densitometry.
引用
收藏
页码:77 / 83
页数:7
相关论文
共 7 条
[1]   MEASUREMENT OF BONE MINERAL IN VIVO - AN IMPROVED METHOD [J].
CAMERONJR. ;
SORENSON, J .
SCIENCE, 1963, 142 (358) :230-+
[2]   LONG-TERM REPRODUCIBILITY OF BONE-MINERAL CONTENT MEASUREMENTS [J].
CHRISTIANSEN, C ;
RODBRO, P .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1977, 37 (04) :321-323
[3]  
LARCOS G, 1991, J NUCL MED, V32, P2101
[4]  
MAZESS RB, 1989, P SOC EXP BIOL MED, V191, P261
[5]   COMPARISON OF DUAL PHOTON AND DUAL ENERGY X-RAY BONE DENSITOMETERS IN A CLINIC SETTING [J].
NELSON, DA ;
BROWN, EB ;
FLYNN, MJ ;
CODY, DD ;
SHAFFER, S .
SKELETAL RADIOLOGY, 1991, 20 (08) :591-595
[6]   DUAL-ENERGY RADIOGRAPHIC ABSORPTIOMETRY FOR BONE DENSITOMETRY - CURRENT STATUS AND PERSPECTIVE [J].
SARTORIS, DJ ;
RESNICK, D .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1989, 152 (02) :241-246
[7]   DUAL-ENERGY X-RAY ABSORPTIOMETRY VERSUS SINGLE PHOTON-ABSORPTIOMETRY OF THE RADIUS [J].
WEINSTEIN, RS ;
NEW, KD ;
SAPPINGTON, LJ .
CALCIFIED TISSUE INTERNATIONAL, 1991, 49 (05) :313-316