VENTRICULAR VOLUME MEASUREMENT FROM A MULTIPLANAR TRANSESOPHAGEAL ULTRASONIC-IMAGING SYSTEM - AN INVITRO STUDY

被引:25
作者
MARTIN, RW
BASHEIN, G
DETMER, PR
MORITZ, WE
机构
[1] UNIV WASHINGTON,CTR BIOENGN,SEATTLE,WA 98195
[2] UNIV WASHINGTON,DEPT ELECT ENGN,SEATTLE,WA 98195
关键词
D O I
10.1109/10.55635
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We have developed a system to assess the feasibility of using multiple transesophageal ultrasonic images to measure left-ventricular volume, an important variable in patient management. The system includes a special transesophageal probe with a micromanipulator for acquiring cardiac images in multiple planes with known interplanar spatial relationship and an off-line processing system to compute the volume. In vitro studies with the probe demonstrated that the distance between two targets in space can be identified within 2 mm (SD = 0.4 mm) for points in the imaging plane 3.4 mm (SD = 0.5 mm) for points not lying in the imaging plane. This gives an average accuracy of ±6.5% for distances greater than 4.5 cm. Comparison of ultrasonic measurements of the volume of water-filled balloons and excised hearts to the volume required to fill them, revealed a correlation coefficient of 0.992, a regression line having a slope of 1.0 and an ordinate intercept at 0.2 ml., and a standard error of the estimate of 8 mL. © 1990 IEEE
引用
收藏
页码:442 / 449
页数:8
相关论文
共 24 条
[21]  
ROSENTHAL MH, 1981, ASA REFRESHER COURSE, V9, P155
[22]   VENTRICULAR PRESSURE-VOLUME DIAGRAM REVISITED [J].
SAGAWA, K .
CIRCULATION RESEARCH, 1978, 43 (05) :677-687
[23]  
Saidman LJ, 1984, MONITORING ANESTHESI
[24]  
Visser C A, 1988, J Cardiothorac Anesth, V2, P74, DOI 10.1016/0888-6296(88)90152-4