LINEAR-RESPONSE FUNCTION FOR FLUORESCENCE-BASED FIBEROPTIC CO2 SENSORS

被引:27
作者
HE, XY [1 ]
RECHNITZ, GA [1 ]
机构
[1] UNIV HAWAII,DEPT CHEM,HAWAII BIOSENSOR LAB,HONOLULU,HI 96822
关键词
D O I
10.1021/ac00109a054
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The theoretical basis and experimental evidence for a Stern-Volmer type relationship using the fluorescence ratio (I-bl/I) in the absence (I-bl) and presence (I) of CO2 as response function for fluorescence-based fiber-optic CO2 sensors is explored. By choosing the right indicator, appropriate bicarbonate concentration, and a thin enough sensing layer, a fiber-optic CO2 sensor with a usable linear range can be easily achieved using I-bl/I as the response function. This allows the possibility of one-point calibration, which cannot be accomplished if the measured fluorescence intensity (I) is used directly as the response function.
引用
收藏
页码:2264 / 2268
页数:5
相关论文
共 18 条
[1]  
Blum L.J., 1991, BIOSENSOR PRINCIPLES
[2]  
DANIELS F, 1961, PHYSICAL CHEM
[3]   COMPARISON OF 2 FIBEROPTIC L-GLUTAMATE BIOSENSORS BASED ON THE DETECTION OF OXYGEN OR CARBON-DIOXIDE, AND THEIR APPLICATION IN COMBINATION WITH FLOW-INJECTION ANALYSIS TO THE DETERMINATION OF GLUTAMATE [J].
DREMEL, BAA ;
SCHMID, RD ;
WOLFBEIS, OS .
ANALYTICA CHIMICA ACTA, 1991, 248 (02) :351-359
[4]  
FORSTER T, 1950, Z ELEKTROCHEM, V54, P531
[5]  
FORSTER T, 1950, Z ELEKTROCHEM, V54, P42
[6]  
HOLLAND JF, 1977, ANAL CHEM, V49, P703
[7]  
LUBBERS DW, 1983, SENSOR ACTUATOR, V4, P641, DOI 10.1016/0250-6874(83)85078-1
[8]   FLUORESCENCE OPTICAL SENSORS FOR CONTINUOUS DETERMINATION OF NEAR-NEUTRAL PH VALUES [J].
OFFENBACHER, H ;
WOLFBEIS, OS ;
FURLINGER, E .
SENSORS AND ACTUATORS, 1986, 9 (01) :73-84
[9]  
Parker C.A., 1968, PHOTOLUMINESCENCE SO
[10]   SOME EXPERIMENTS WITH SPECTROFLUORIMETERS AND FILTER FLUORIMETERS [J].
PARKER, CA ;
BARNES, WJ .
ANALYST, 1957, 82 (978) :606-&