IMPROVED MASS ACCURACY IN MATRIX-ASSISTED LASER DESORPTION/IONIZATION TIME-OF-FLIGHT MASS-SPECTROMETRY OF PEPTIDES

被引:137
作者
VORM, O
MANN, M
机构
[1] EUROPEAN MOLEC BIOL LAB,PROT & PEPTIDE GRP,D-69012 HEIDELBERG,GERMANY
[2] ODENSE UNIV,DEPT BIOL MOLEC,ODENSE,DENMARK
关键词
D O I
10.1016/1044-0305(94)80013-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
One problem of matrix-assisted laser desorption ionization coupled to time-of-flight mass spectrometry is the moderate mass accuracy that typically can be obtained in routine applications. Here we report improved mass accuracy for peptides, even when low amounts and complex peptide mixtures are used. A new procedure for preparing matrix surfaces is used, and there is no need to mix the matrix with the sample or to add internal standards. Examples are shown with a mass accuracy better than 50 ppm in a peptide mixture. Peptide mapping as well as sequencing by creating ''ragged ends'' or ''ladder sequencing'' should benefit especially from the improved mass accuracy.