CALIBRATION OF SILICON PIN PHOTODIODE FOR MEASURING INTENSITY OF 7-SIMILAR-TO-40 KEV PHOTONS

被引:7
作者
BAN, S
HIRAYAMA, H
NAMITO, Y
TANAKA, S
NAKASHIMA, H
NAKANE, Y
NARIYAMA, N
机构
[1] JAPAN ATOM ENERGY RES INST, TOKAI, IBARAKI 31911, JAPAN
[2] SHIP RES INST, MITAKA, TOKYO 181, JAPAN
关键词
SILICON PIN PHOTODIODES; CALIBRATION; CALIBRATION STANDARD; PHOTONS; X-RADIATION; SYNCHROTRON RADIATION; KEV RANGE 01-10; KEV RANGE 10-100; MEASURED VALUE; ACCURACY;
D O I
10.3327/jnst.31.163
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Silicon PIN photodiodes -ere used to measure the absolute intensity of monoenergetic photons in a synchrotron radiation beam line at 7-40 keV region. Below 20 keV, the intensity was measured with an accuracy of 5%. Above this energy, the accuracy was about 20% because the photocurrent in the sensitive area was not uniform due to backscattered radiation from backing.
引用
收藏
页码:163 / 168
页数:6
相关论文
共 14 条
[1]   DESIGN OF BEAMLINE 14 FOR THE PF VERTICAL WIGGLER AND ITS OPERATION [J].
ANDO, M ;
SATOW, Y ;
KAWATA, H ;
ISHIKAWA, T ;
SPIEKER, P ;
SUZUKI, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 246 (1-3) :144-148
[2]   MEASUREMENT OF THE PHOTON ENERGY-ABSORPTION COEFFICIENT FOR AIR, NITROGEN AND ARGON AT 30-KEV [J].
BAN, S ;
HIRAYAMA, H ;
NAMITO, Y ;
TANAKA, S ;
NAKASHIMA, H ;
NAKANE, Y ;
YOSHIZAWA, M ;
NARIYAMA, N .
APPLIED RADIATION AND ISOTOPES, 1993, 44 (04) :769-772
[3]   SILICON PHOTODIODE DETECTOR FOR FLUORESCENCE EXAFS [J].
BOULDIN, CE ;
FORMAN, RA ;
BELL, MI .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1987, 58 (10) :1891-1894
[4]   MEASUREMENT OF LOW ENERGY X-RAYS .V. TOTAL ABSORPTION SILICON DEVICES [J].
GREENING, JR ;
RANDLE, KJ ;
REDPATH, AT .
PHYSICS IN MEDICINE AND BIOLOGY, 1969, 14 (01) :55-&
[5]   MEASUREMENT OF LOW ENERGY X-RAYS .I. GENERAL CONSIDERATIONS [J].
GREENING, JR ;
RANDLE, KJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1968, 13 (02) :159-&
[6]  
GREENING JR, 1970, ICRU17 REP
[7]   PHOTON MASS ATTENUATION AND ENERGY-ABSORPTION COEFFICIENTS FROM 1 KEV TO 20 MEV [J].
HUBBELL, JH .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (11) :1269-1290
[8]   PIN DIODES AS DETECTORS IN THE ENERGY REGION 500 EV-10 KEV [J].
JACH, T ;
COWAN, PL .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 208 (1-3) :423-425
[9]   PIN DIODE DETECTORS FOR SYNCHROTRON X-RAYS [J].
KIRKLAND, JP ;
JACH, T ;
NEISER, RA ;
BOULDIN, CE .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 266 (1-3) :602-607
[10]   CALIBRATION OF SEMICONDUCTOR PHOTODIODES AS SOFT-X-RAY DETECTORS [J].
KRUMREY, M ;
TEGELER, E ;
THORNAGEL, R ;
ULM, G .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (07) :2291-2294