PHOTOPOLARIMETER BASED ON PLANAR GRATING DIFFRACTION

被引:31
作者
AZZAM, RMA
GIARDINA, KA
机构
[1] Department of Electrical Engineering, University of New Orleans, Lakefront, New Orleans, LA
来源
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION | 1993年 / 10卷 / 06期
关键词
D O I
10.1364/JOSAA.10.001190
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A division-of-amplitude photopolarimeter (DOAP) is described that employs a diffraction grating in the conventional spectrometer orientation with the grating grooves normal to the plane of incidence. Four coplanar diffracted orders are used for polarimetric analysis to determine all four Stokes parameters of incident light simultaneously and virtually instantaneously (with the speed being determined solely by the photodetectors and their associated electronics); a fifth order is used for alignment by autocollimation or by use of a position-sensing quadrant detector. To sensitize the instrument for the +45-degrees and -45-degrees azimuths of incident linearly polarized light and for the handedness of incident circular polarization (i.e., for the third and fourth Stokes parameters), we insert two linear polarizers in two diffracted orders with their transmission axes inclined at appropriate angles with respect to the plane of incidence. The calibration and testing of an instrument of this type that uses an Al-coated 600-groove/mm holographic grating at 632.8-nm wavelength are reported as an example.
引用
收藏
页码:1190 / 1196
页数:7
相关论文
共 38 条
[1]  
[Anonymous], 1974, PLANETS STARS NEBULA
[2]  
AZZAM RAM, 1992, DIFFRACTIVE OPTICS D, V9, P69
[3]   PRECISION ANALYSIS AND LOW-LIGHT-LEVEL MEASUREMENTS USING A PROTOTYPE 4-DETECTOR PHOTOPOLARIMETER (FDP) [J].
AZZAM, RMA ;
LOPEZ, AG .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (08) :2063-2068
[4]  
AZZAM RMA, 1979, OPTIK, V52, P253
[5]   INSTRUMENT MATRIX OF THE 4-DETECTOR PHOTOPOLARIMETER - PHYSICAL MEANING OF ITS ROWS AND COLUMNS AND CONSTRAINTS ON ITS ELEMENTS [J].
AZZAM, RMA .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1990, 7 (01) :87-91
[6]   DIVISION-OF-AMPLITUDE PHOTOPOLARIMETER BASED ON CONICAL DIFFRACTION FROM A METALLIC GRATING [J].
AZZAM, RMA .
APPLIED OPTICS, 1992, 31 (19) :3574-3576
[7]   CONSTRUCTION, CALIBRATION, AND TESTING OF A 4-DETECTOR PHOTOPOLARIMETER [J].
AZZAM, RMA ;
MASETTI, E ;
ELMINYAWI, IM ;
GROSZ, FG .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (01) :84-88
[8]   BEAM-SPLITTERS FOR THE DIVISION-OF-AMPLITUDE PHOTOPOLARIMETER [J].
AZZAM, RMA .
OPTICA ACTA, 1985, 32 (11) :1407-1412
[9]   ACCURATE CALIBRATION OF THE 4-DETECTOR PHOTOPOLARIMETER WITH IMPERFECT POLARIZING OPTICAL-ELEMENTS [J].
AZZAM, RMA ;
LOPEZ, AG .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1989, 6 (10) :1513-1521
[10]   ARRANGEMENT OF 4 PHOTODETECTORS FOR MEASURING THE STATE OF POLARIZATION OF LIGHT [J].
AZZAM, RMA .
OPTICS LETTERS, 1985, 10 (07) :309-311