MULTICENTER CLINICAL-EVALUATION OF A NEW SSIR PACEMAKER

被引:4
作者
BONGIORNI, MG
SOLDATI, E
ARENA, G
DESIMONE, L
CAPUCCI, A
GALLI, R
PARLAPIANO, M
CAZZIN, R
MORACCHINI, P
LEONARDI, C
ZARDO, F
OMETTO, R
VINCENZI, M
CHARLES, R
MAKIN, C
BIAGINI, A
机构
[1] UNIV BOLOGNA,INST CARDIOSURG,I-40126 BOLOGNA,ITALY
[2] UNIV BOLOGNA,INST CARDIOL,I-40126 BOLOGNA,ITALY
[3] CARDIOL DEPT,MESTRE,ITALY
[4] CARDIOL DEPT,MODENA,ITALY
[5] CARDIOL DEPT,PORDENONE,ITALY
[6] CARDIOL DEPT,VICENZA,ITALY
[7] CARDIOTHORAC CTR,LIVERPOOL,ENGLAND
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1992年 / 15卷 / 11期
关键词
RATE RESPONSIVE PACING; ACCELERATION SENSOR; RATE ADAPTIVE PACEMAKER;
D O I
10.1111/j.1540-8159.1992.tb02971.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A multicenter clinical evaluation of Sorin Swing 100, a new SSIR pacemaker with a gravimetric sensor, was performed by seven different centers enrolling a total of 89 patients, 56 men and 33 women, mean age 73.1 years, for pacemaker implantion (73 patients) or pacemaker replacement (1 6 patients). Pacing mode was VVIR in 73 patients and AAIR in 16. The behavior of pacing rate was evaluated 3 months after the implant by performing a 24-hour Holter monitor, an exercise stress test, and tests for the assessment of mechanical external interference (MEI). A physiological behavior of the paced rate was always observed during Holter monitoring. In 52 completely paced patients mean diurnal, nocturnal, and maximal heart rate were, respectively, 74.9 +/- 5.7 ppm, 58.1 +/- 5.8 ppm, and 113.4 +/- 12.7 ppm; a paced rate exceeding 100 ppm was reached on the average 5.6 times/Holter monitor. In all but two patients the sleep rate (55 ppm) was reached during the night or long resting time. During exercise stress test a direct correlation between the increase in pacing rate and the increase in workload was observed; the mean maximal heart rate reached in 49 completely paced patients was, respectively, 102.8 +/- 9 ppm in 17 patients who accomplished stage 1, 116.2 +/- 13.6 ppm in 28 patients who accomplished stage 2, and 133 +/- 6.7 ppm in 10 patients who accomplished stage 3 of the Bruce protocol. MEI testing never increased the pacing rate over the noise rate (10 ppm over the basic rate). In only seven patients the results obtained suggested to change the nominal set up of the pacemaker. Our experience clearly indicates that Swing 1 00 is an effective, reliable, and easy to use SSIR pacemaker. The availability of the sleep rate allows a more physiological pattern of pacing rate and can lead to significant energy saving.
引用
收藏
页码:1798 / 1803
页数:6
相关论文
共 12 条
[11]   RATE RESPONSIVE PACING AND EXERCISE CAPACITY - A COMMENT [J].
NORDLANDER, R ;
HEDMAN, A ;
PEHRSSON, SK .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (05) :749-751
[12]   PHYSICAL MOVEMENT SENSITIVE PACING - COMPARISON OF 2 ACTIVITY TRIGGERED PACING SYSTEMS [J].
STANGL, K ;
WIRTZFELD, A ;
LOCHSCHMIDT, O ;
BASLER, B ;
MITTNACHT, A .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (01) :102-110