EVALUATION OF THE HEMODYNAMIC PERFORMANCE OF SMALL CARBOMEDICS AORTIC PROSTHESES USING DOBUTAMINE-STRESS DOPPLER-ECHOCARDIOGRAPHY

被引:33
作者
IZZAT, MB [1 ]
BIRDI, I [1 ]
WILDE, P [1 ]
BRYAN, AJ [1 ]
ANGELINI, GD [1 ]
机构
[1] UNIV BRISTOL,BRISTOL ROYAL INFIRM,DEPT CARDIAC RADIOL,BRISTOL BS2 8HW,AVON,ENGLAND
关键词
D O I
10.1016/0003-4975(95)00462-T
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. The well-known correlation between prosthetic valve orifice area and transvalvular gradients has raised concerns about the presence of significant residual gradients when the size of the prosthesis that can be implanted is limited by the presence of a small aortic annulus. Methods. Dobutamine-stress Doppler echocardiography was used to evaluate the hemodynamic performance of small CarboMedics aortic prostheses (19 mm and 21 mm) in 18 patients (16 women; mean age, 64 years) who had undergone aortic valve replacement 23.5 +/- 19 months (standard deviation) previously. Dobutamine infusion was started at a rate of 5 mu g . kg(-1). min(-2) and increased to 10 and 20 mu g . kg(-1). min(-2) at 15-minute intervals. Pulsed and continuous wave Doppler studies were performed at rest and at the end of each stage. Effective orifice area, performance index, and discharge coefficient of both valves were calculated, and peak and mean velocity and pressure drop across the prostheses were measured. Results. Heart rate and cardiac output increased by 74% and 94%, respectively, and mean arterial blood pressure decreased by 9% at maximum stress. Effective orifice area, discharge coefficient, and performance index were comparable in both valve sizes at rest and maximum stress. Also, there was no significant difference in mean transvalvular pressure drop (gradient) for 19-mm and 21-mm prostheses at rest (8.1 +/- 8.4 and 4.8 +/- 3.8 mm Hg) or maximum stress (15.1 +/- 14.2 and 8.8 +/- 5.8 mm Hg, respectively). No significant correlation could be demonstrated between transvalvular pressure drop and patient's body surface area. Conclusions. These data show that 19-mm and 21-mm CarboMedics aortic prostheses exhibit equally favorable hemodynamic performance with minimal pressure gradient, both at rest and under stress conditions.
引用
收藏
页码:1048 / 1052
页数:5
相关论文
共 22 条
[1]  
Arom K V, 1994, J Heart Valve Dis, V3, P531
[2]   DISCREPANCIES BETWEEN DOPPLER AND CATHETER GRADIENTS IN AORTIC PROSTHETIC VALVES INVITRO - A MANIFESTATION OF LOCALIZED GRADIENTS AND PRESSURE RECOVERY [J].
BAUMGARTNER, H ;
KHAN, S ;
DEROBERTIS, M ;
CZER, L ;
MAURER, G .
CIRCULATION, 1990, 82 (04) :1467-1475
[3]   HEMODYNAMIC CHANGES AFTER VALVE-REPLACEMENT WITH STARR-EDWARDS PROSTHESES [J].
BRISTOW, JD ;
KREMKAU, EL .
AMERICAN JOURNAL OF CARDIOLOGY, 1975, 35 (05) :716-724
[4]   DOPPLER ECHOCARDIOGRAPHIC ASSESSMENT OF THE ST-JUDE MEDICAL PROSTHETIC VALVE IN THE AORTIC POSITION USING THE CONTINUITY EQUATION [J].
CHAFIZADEH, ER ;
ZOGHBI, WA .
CIRCULATION, 1991, 83 (01) :213-223
[5]   ECHOCARDIOGRAPHIC DESCRIPTION OF THE CARBOMEDICS BILEAFLET PROSTHETIC HEART-VALVE [J].
CHAMBERS, J ;
CROSS, J ;
DEVERALL, P ;
SOWTON, E .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (02) :398-405
[6]  
CHAMBERS J, 1994, BRIT HEART J, V71, P6
[7]  
DEPAULIS R, 1994, J THORAC CARDIOV SUR, V108, P57
[8]   VALIDATION AND APPLICATIONS OF INDEXED AORTIC PROSTHETIC VALVE AREAS CALCULATED BY DOPPLER ECHOCARDIOGRAPHY [J].
DUMESNIL, JG ;
HONOS, GN ;
LEMIEUX, M ;
BEAUCHEMIN, J .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) :637-643
[9]  
Fisher J, 1994, J Heart Valve Dis, V3, P75
[10]   HEMODYNAMIC EVALUATION OF THE CARBOMEDICS PROSTHETIC HEART-VALVE IN THE AORTIC POSITION - COMPARISON OF NONINVASIVE AND INVASIVE TECHNIQUES [J].
IHLEN, H ;
MOLSTAD, P ;
SIMONSEN, S ;
VATNE, K ;
OVRUM, E ;
GEIRAN, O ;
LAAKE, P ;
FROYSAKER, T .
AMERICAN HEART JOURNAL, 1992, 123 (01) :151-159