DEVELOPMENT OF AN ENZYME-BASED BIOSENSOR FOR ATRAZINE DETECTION

被引:68
作者
MCARDLE, FA [1 ]
PERSAUD, KC [1 ]
机构
[1] UMIST,DEPT INSTRUMENTAT & ANALYT SCI,MANCHESTER M60 1QD,ENGLAND
关键词
BIOSENSOR; ATRAZINE; TYROSINASE; ENZYME IMMOBILIZATION; AMPEROMETRY;
D O I
10.1039/an9931800419
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An amperometric biosensor for detection of the photosynthetic inhibiting herbicide atrazine is described. The sensor, incorporating the enzyme tyrosinase, was sensitive to a wide range of di- and triphenols. It displayed a fast response, with 95% of the steady-state current being obtained within 25 s, and had a half-life of 8 d but retained some activity for up to 20 d. This biosensor displayed a notable decrease in response in the presence of atrazine. Atrazine inhibition of the biosensor response was found to be reversible. The electrode construction consisted of tyrosinase cross-linked on a poly(pyrrole) coated gold electrode surface. The inhibition of enzyme activity in the presence of atrazine was measured amperometrically. Atrazine concentrations of 5 x 10(-6) mol dm-3 were measured in a hydrodynamic cell and of 1 X 10(-5) mol dm-3 in a flow-through system.
引用
收藏
页码:419 / 423
页数:5
相关论文
共 19 条