NITRATE ION-SELECTIVE ELECTRODE AS REFERENCE ELECTRODE FOR FLOW-INJECTION ANALYSIS

被引:5
作者
ALONSOCHAMARRO, J
BARTROLI, J
JIMENEZ, C
机构
[1] Facultat de Ciències, Unitat de Química Analítica, Department de Química, 08193 Bellaterra, Barcelona
关键词
FLOW SYSTEM; POTENTIOMETRY; ION-SELECTIVE ELECTRODES; ION-SENSITIVE FIELD-EFFECT TRANSISTORS; PH; NITRATE;
D O I
10.1016/0003-2670(92)80222-S
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The development of a flow-injection system for pH determination using a pH ion-sensitive field-effect transistor in conjunction with a tubular flow-through nitrate ion-selective etectrode (ISE) as a reference electrode is described. With this system, the linear working range extends from pH 1 to 13 with a sensitivity of 56 mV per pH unit and a relative standard deviation of 0.4%. A stable baseline and good response are obtained by placing the tubular ISE in a closed branch of the main channel.
引用
收藏
页码:419 / 423
页数:5
相关论文
共 15 条
[1]   FLOW-THROUGH SANDWICH PVC MATRIX MEMBRANE-ELECTRODE FOR FLOW-INJECTION ANALYSIS [J].
ALEGRET, S ;
ALONSO, J ;
BARTROLI, J ;
LIMA, JLFC ;
MACHADO, AASC ;
PAULIS, JM .
ANALYTICAL LETTERS PART A-CHEMICAL ANALYSIS, 1985, 18 (18) :2291-2303
[2]   FLOW-THROUGH PH-ISFET AS DETECTOR IN THE DETERMINATION OF AMMONIA [J].
ALEGRET, S ;
ALONSO, J ;
BARTROLI, J ;
DELVALLE, M ;
JAFFREZICRENAULT, N ;
DUVAULTHERRERA, Y .
ANALYTICA CHIMICA ACTA, 1990, 231 (01) :53-58
[3]   FLOW-THROUGH TUBULAR PVC MATRIX MEMBRANE-ELECTRODE WITHOUT INNER REFERENCE SOLUTION FOR FLOW-INJECTION ANALYSIS [J].
ALEGRET, S ;
ALONSO, J ;
BARTROLI, J ;
PAULIS, JM ;
LIMA, JLFC ;
MACHADO, AASC .
ANALYTICA CHIMICA ACTA, 1984, 164 (OCT) :147-152
[4]   FLOW-THROUGH PH-ISFET AS DETECTOR IN AUTOMATED DETERMINATIONS [J].
ALEGRET, S ;
BARTROLI, J ;
JIMENEZ, C ;
DELVALLE, M ;
DOMINGUEZ, C ;
CABRUJA, E ;
MERLOS, A .
ELECTROANALYSIS, 1991, 3 (4-5) :349-354
[5]  
ALEGRET S, 1991, IN PRESS SENSORS ACT
[6]  
ALEGRET S, 1986, 2ND P INT M CHEM SEN, P751
[7]   SEQUENTIAL FLOW-INJECTION DETERMINATIONS OF CALCIUM AND MAGNESIUM IN WATERS [J].
ALONSO, J ;
BARTROLI, J ;
LIMA, JLFC ;
MACHADO, AASC .
ANALYTICA CHIMICA ACTA, 1986, 179 :503-508
[8]   THE OPERATION OF AN ISFET AS AN ELECTRONIC DEVICE [J].
BERGVELD, P .
SENSORS AND ACTUATORS, 1981, 1 (01) :17-29
[9]  
CANE C, 1991, TRANSDUCERS, V91, P225
[10]  