QUANTITATIVE TRANSMISSION-X-RAY MICROANALYSIS OF IONIC COMPOUNDS

被引:147
作者
VANCAPPELLEN, E [1 ]
DOUKHAN, JC [1 ]
机构
[1] UNIV LILLE 1,STRUCT & PROPRIETES ETAT SOLIDE LAB,URA 234,F-59655 VILLENEUVE DASCQ,FRANCE
关键词
Electroneutrality - Ionic compounds - Quantitative transmission x ray microanalysis - Ultra thin window x ray detector - X ray take off angle;
D O I
10.1016/0304-3991(94)90047-7
中图分类号
TH742 [显微镜];
学科分类号
摘要
A new absorption correction applicable to EDX spectra of ionic compounds (oxides, ceramics, minerals,...) is presented. The method enables the accurate quantification of oxygen and nitrogen, and does not require parameters such as specimen thickness, X-ray take-off angle and specimen density. Physically this new correction procedure is based on the sole principle of electroneutrality of the specimen: this means that the sum of all anions and cations times their respective valence states must cancel out. An additional benefit of this method is the possibility to calculate the average thickness of the analysed area if, after the quantitative analysis, numbers for. the specimen density and X-ray take-off angle are available.
引用
收藏
页码:343 / 349
页数:7
相关论文
共 3 条
[1]  
CLIFF G, 1972, 5TH P EUR C EL MICR, P140
[2]  
TIXIER R, 1969, 5TH P INT C XRAY OPT, P180
[3]   THE PARAMETERLESS CORRECTION METHOD IN X-RAY-MICROANALYSIS [J].
VANCAPPELLEN, E .
MICROSCOPY MICROANALYSIS MICROSTRUCTURES, 1990, 1 (01) :1-22