THE RELATIONSHIP OF EFFECTIVE DOSE TO PERSONNEL AND MONITOR READING FOR SIMULATED FLUOROSCOPIC IRRADIATION CONDITIONS

被引:57
作者
FAULKNER, K
MARSHALL, NW
机构
[1] Regional Medical Physics Department, Newcastle General Hospital, Newcastle upon Tyne, Westgate Road
来源
HEALTH PHYSICS | 1993年 / 64卷 / 05期
关键词
X-RAY IMAGING; EXPOSURE; OCCUPATIONAL; DOSIMETRY; PERSONNEL; DIAGNOSTIC RADIOLOGY;
D O I
10.1097/00004032-199305000-00007
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The exposure of staff during fluoroscopic procedures was simulated for overcouch x-ray tube/undercouch image intensifier and undercouch x-ray tube/overcouch image intensifier geometries. A Rando phantom with film badge dosimeters attached to the skin surface at seven commonly used monitoring sites and loaded with lithium fluoride thermoluminescent dosimeters was irradiated for an extended period in the vicinity of a patient couch. Scattered radiation generated from the irradiation of an anthropomorphic phantom using primary radiation in the range of 70 kVp-110 kVp was used. The radiation dose to organs which were shielded by a lead apron was estimated from the unattenuated organ dose readings by applying an experimentally determined scattered radiation transmission factor. The ratio of effective dose to film badge reading was obtained for a range of irradiation conditions and lead apron thicknesses. For most irradiation conditions studied, a dosimeter worn above the lead apron will significantly overestimate effective dose by a factor of between 2 and 60, depending on the irradiation conditions. A dosimeter worn under the apron at either waist or chest level, will generally yield a closer (although usually an underestimate of) effective dose, typically within a factor of 7 for the most common lead apron thicknesses and irradiation conditions. No single dosimeter can accurately monitor effective dose for all irradiation conditions in fluoroscopy.
引用
收藏
页码:502 / 508
页数:7
相关论文
共 13 条
[1]  
BRATEMAN L, 1989, HEALTH PHYS, V56, P567
[2]  
BUSHONG SC, 1989, HEALTH PHYS, V56, P565
[3]   ESTIMATION OF EFFECTIVE DOSE EQUIVALENT TO STAFF IN DIAGNOSTIC-RADIOLOGY [J].
FAULKNER, K ;
HARRISON, RM .
PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (01) :83-91
[4]  
HUDA W, 1984, HEALTH PHYS, V47, P463
[5]   MONITORING OF STAFF WEARING LEAD-RUBBER APRONS [J].
JONES, JR .
BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (706) :1051-1052
[6]   THE DEPENDENCE OF THE SCATTERED RADIATION-DOSE TO PERSONNEL ON TECHNIQUE FACTORS IN DIAGNOSTIC-RADIOLOGY [J].
MARSHALL, NW ;
FAULKNER, K .
BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (769) :44-49
[7]  
MEINHOLD CB, 1989, HEALTH PHYS, V56, P570
[8]   BROAD-BEAM TRANSMISSION DATA IN LEAD FOR SCATTERED RADIATION PRODUCED AT DIAGNOSTIC ENERGIES [J].
RAWLINGS, DJ ;
FAULKNER, K ;
HARRISON, RM .
BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (757) :69-71
[9]  
WEBSTER EW, 1989, HEALTH PHYS, V56, P568
[10]  
1988, 1988 US NAT SCI COMM