EFFECTS OF GOLD AND SILVER BACKINGS ON THE DOSE-RATE AROUND AN I-125 SEED

被引:23
作者
CYGLER, J [1 ]
SZANTO, J [1 ]
SOUBRA, M [1 ]
ROGERS, DWO [1 ]
机构
[1] NATL RES COUNCIL CANADA, DIV PHYS, IONIZING RADIAT STAND, OTTAWA K1A 0R6, ONTARIO, CANADA
关键词
125I seeds; brachytherapy; dose rate; fluorescent x-rays; gold backing;
D O I
10.1118/1.596588
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Measurements of the effect of either gold or silver backing on the dose rate around an 125I seed were performed using a Therados RFA7 dosimetry system and a small diode detector which was 2.5 mm in diameter and 0.06 mm thick. It was found that the presence of the gold or silver backing modifies the diode response on the side of the 125I seed away from the backing. The effect depends on the backing material and the distance from the seed. There is a small increase close to the gold backing but a decrease further away. This decrease at distances greater than 10 mm from the seed is uniformly 10%, the same as found when the seed is backed by air. There is an increase of up to 25% observed with silver backing the seed and this increase remains significant more than 30 mm from the seed. When the response increases, the results are hard to interpret quantitatively because of variations in the diode response per unit dose with photon energy and extreme sensitivity to geometric changes. Nonetheless, except for the increase at close distances with the gold, the results are in agreement with EGS4 Monte Carlo photon transport simulations which are for a simplified geometry and account for x-ray fluorescence from the K-shell. Furthermore, the increase in the gold-backed case is qualitatively explained by Williamson’s Monte Carlo calculations which take into account the L-shell fluorescent x-rays from gold. © 1990, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:172 / 178
页数:7
相关论文
共 22 条
[11]   BACKSCATTER MEASUREMENTS FROM A SINGLE SEED OF I125 FOR OPHTHALMIC PLAQUE DOSIMETRY [J].
LUXTON, G ;
ASTRAHAN, MA ;
PETROVICH, Z .
MEDICAL PHYSICS, 1988, 15 (03) :397-400
[12]   A COMPARISON OF SOLID PHANTOMS WITH WATER FOR DOSIMETRY OF I125 BRACHYTHERAPY SOURCES [J].
MEIGOONI, AS ;
MELI, JA ;
NATH, R .
MEDICAL PHYSICS, 1988, 15 (05) :695-701
[13]   EPIPERICARDIAL FAT PAD - CT FINDINGS [J].
PALING, MR ;
WILLIAMSON, BRJ .
RADIOLOGY, 1987, 165 (02) :335-339
[14]   SELECTIVE SHIELDING OF A PARA-SI DETECTOR FOR QUALITY INDEPENDENCE [J].
RIKNER, G ;
GRUSELL, E .
ACTA RADIOLOGICA ONCOLOGY, 1985, 24 (01) :65-69
[15]   THE DOSIMETRY OF I-125 SEED EYE PLAQUES [J].
WEAVER, KA .
MEDICAL PHYSICS, 1986, 13 (01) :78-83
[16]   RESPONSE OF LIF POWDER TO I-125 PHOTONS [J].
WEAVER, KA .
MEDICAL PHYSICS, 1984, 11 (06) :850-854
[17]  
WILLIAMSON JF, 1985, MED PHYS, V12, P676
[18]   MONTE-CARLO EVALUATION OF SPECIFIC DOSE CONSTANTS IN WATER FOR I125 SEEDS [J].
WILLIAMSON, JF .
MEDICAL PHYSICS, 1988, 15 (05) :686-694
[19]  
WILLIAMSON JF, 1988, COMMUNICATION JUL
[20]   EFFECT OF GOLD SHIELDING ON THE DOSIMETRY OF AN I125 SEED AT CLOSE RANGE [J].
WU, A ;
STERNICK, ES ;
MUISE, DJ .
MEDICAL PHYSICS, 1988, 15 (04) :627-628