PREVENTION OF FRACTURE AT THE DISTAL LOCKING SITE OF THE GAMMA-NAIL

被引:52
作者
LACROIX, H
ARWERT, H
SNIJDERS, CJ
FONTIJNE, WPJ
机构
[1] ERASMUS UNIV ROTTERDAM,DEPT BIOMED PHYS & MED TECHNOL,3015 SE ROTTERDAM,NETHERLANDS
[2] UNIV HOSP DIJKZIGT,DEPT ORTHOPAED SURG,3015 GD ROTTERDAM,NETHERLANDS
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1995年 / 77B卷 / 02期
关键词
D O I
10.1302/0301-620X.77B2.7706346
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
To investigate the origin of fractures at the distal locking site of the Gamma nail, we loaded ten paired human cadaver femora fixed with a Gamma nail in torsion until they fractured. When an aml was hammered in to start the hole for distal locking a fissure appeared in the lateral cortex of all the femora, and the mean torsional load to create a fracture was reduced by 57.8% compared with that in a control group in which the distal locking hole had been started with a centre drill. When an additional drill hole was made, the mean failure load in torsion decreased by 35.7%. We strongly recommend that an awl should not be used at the distal locking site of the Gamma nail; we recommend the use of a centre drill. Additional drill holes should be avoided because they act as stress raisers.
引用
收藏
页码:274 / 276
页数:3
相关论文
共 11 条
[1]
BANNISTER GC, 1983, J BONE JOINT SURG BR, V65, P218
[2]
BANNISTER GC, 1990, CLIN ORTHOP RELAT R, V254, P242
[3]
BORIANI S, 1991, ORTHOPEDICS, V14, P1307
[4]
INTERTROCHANTERIC FEMORAL FRACTURES - MECHANICAL FAILURE AFTER INTERNAL-FIXATION [J].
DAVIS, TRC ;
SHER, JL ;
HORSMAN, A ;
SIMPSON, M ;
PORTER, BB ;
CHECKETTS, RG .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (01) :26-31
[5]
TROCHANTERIC FRACTURES OF THE FEMUR - A RANDOMIZED PROSPECTIVE TRIAL COMPARING THE JEWETT NAIL-PLATE WITH THE DYNAMIC HIP SCREW [J].
ESSER, MP ;
KASSAB, JY ;
JONES, DHA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1986, 68 (04) :557-560
[6]
THE GAMMA-NAIL FOR PERITROCHANTERIC FRACTURES [J].
HALDER, SC .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (03) :340-344
[7]
TREATMENT OF INTERTROCHANTERIC FRACTURES OF THE FEMUR - A COMPARISON OF THE RICHARDS SCREW-PLATE WITH THE JEWETT NAIL-PLATE [J].
HEYSEMOORE, GH ;
MACEACHERN, AG ;
EVANS, DCJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1983, 65 (03) :262-267
[8]
GAMMA-NAILS AND DYNAMIC HIP SCREWS FOR PERITROCHANTERIC FRACTURES - A RANDOMIZED PROSPECTIVE-STUDY IN ELDERLY PATIENTS [J].
LEUNG, KS ;
SO, WS ;
SHEN, WY ;
HUI, PW .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (03) :345-351
[9]
EARLY EXPERIENCE WITH THE GAMMA INTERLOCKING NAIL FOR PERITROCHANTERIC FRACTURES OF THE PROXIMAL FEMUR [J].
LINDSEY, RW ;
TEAL, P ;
PROBE, RA ;
RHOADS, D ;
DAVENPORT, S ;
SCHAUDER, K .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1991, 31 (12) :1649-1658
[10]
BONE WEAKNESS AFTER THE REMOVAL OF PLATES AND SCREWS - CORTICAL ATROPHY OR SCREW HOLES [J].
ROSSON, J ;
EGAN, J ;
SHEARER, J ;
MONRO, P .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (02) :283-286