ROOM-TEMPERATURE PHOSPHORESCENCE OPTOSENSOR FOR TETRACYCLINES

被引:36
作者
ALAVAMORENO, F [1 ]
DIAZGARCIA, ME [1 ]
SANZMEDEL, A [1 ]
机构
[1] UNIV OVIEDO,DEPT PHYS & ANALYT CHEM,C JULIAN CLAVERIA 8,E-33006 OVIEDO,SPAIN
关键词
FLOW INJECTION; PHOSPHORIMETRY; SENSORS; OPTOSENSORS; TETRACYCLINES;
D O I
10.1016/0003-2670(93)85025-F
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow-through optosensor for tetracyclines based on the tetracycline-europium chelate room temperature phosphorescence energy transfer is proposed. The sensor is developed in conjunction with a flow-injection analysis system and is based on the transient immobilization on a non-ionic resin (packed in a flow-through cell) of the tetracycline-europium chelate. The analytical performance characteristics of the proposed sensor for semiautomated analysis and control of very low levels of tetracycline were as follows: the detection limits for tetracycline, oxytetracycline and chlortetracycline were 0.25, 0.30 and 0.40 ng ml-1, respectively, with a relative standard deviation of 1% for determination of 0.24 mug ml-1 of each antibiotic (n = 10). Most of the common metal ions in biological samples did not interfere, except Fe(III) which caused serious interference and should be masked with 1,10-phenanthroline. The recommended method has been successfully tested for determination of tetracyclines in clinical samples (urine and pharmaceutical preparations).
引用
收藏
页码:637 / 644
页数:8
相关论文
共 18 条