A COMPARISON OF THE PERFORMANCE OF 20 PULSE OXIMETERS UNDER CONDITIONS OF POOR PERFUSION

被引:81
作者
CLAYTON, DG
WEBB, RK
RALSTON, AC
DUTHIE, D
RUNCIMAN, WB
机构
[1] Head of Department, University of Adelaide, Australian Patient Safety Foundation, Adelaide, South Australia, 5001
关键词
EQUIPMENT; PULSE OXIMETRY; MEASUREMENT;
D O I
10.1111/j.1365-2044.1991.tb09303.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The performance of 20 pulse oximeters with finger probes was evaluated by comparison of their readings with directly measured arterial blood oxygen saturations. The samples were taken from patients who had undergone cardiac surgery under hypothermic cardiopulmonary bypass and had poor peripheral perfusion. The mean difference (bias, accuracy), standard deviation (precision) and drop‐out rate for each pulse oximeter was determined. An overall ranking of performance of each pulse oximeter was calculated using five criteria (accuracy, precision, number of readings within 3% of standard, percentage of readings given within 3% of standard, expectedoverread limit in 95% of cases). Two pulse oximeters achieved a combination of accuracy and precision such that 95% of measurements would be expected to be within 4% of the co‐oximeter value; these two also had the lowest dropout rate. Copyright © 1991, Wiley Blackwell. All rights reserved
引用
收藏
页码:3 / 10
页数:8
相关论文
共 15 条
[11]  
TREMPER KK, 1989, ANESTHESIOLOGY, V70, P98, DOI 10.1097/00000542-198901000-00019
[12]  
TREMPER KK, 1985, ANESTHESIOLOGY, V63, P175
[13]  
WEBB RK, 1989, ANAESTH INTENS CARE, V17, P238
[14]   COMPARISON OF PULSE OXIMETERS - EFFECTS OF VASOCONSTRICTION AND VENOUS ENGORGEMENT [J].
WILKINS, CJ ;
MOORES, M ;
HANNING, CD .
BRITISH JOURNAL OF ANAESTHESIA, 1989, 62 (04) :439-444
[15]  
1989, HLTH DEVICES, V18, P183