BACTERIAL RETENTION PROPERTIES OF HEAT AND MOISTURE EXCHANGE FILTERS

被引:33
作者
LEE, MG
FORD, JL
HUNT, PB
IRELAND, DS
SWANSON, PW
机构
[1] LIVERPOOL JOHN MOORES UNIV,SCH PHARM,LIVERPOOL L3 3AF,ENGLAND
[2] WHISTON HOSP,PHARMACEUT MICROBIOL LAB,PRESCOT L35 5DR,MERSEYSIDE,ENGLAND
[3] COUNTESS CHESTER HOSP,QC LAB,CHESTER CH1 2BA,ENGLAND
关键词
BACTERIA; FILTRATION; EQUIPMENT; HEAT AND MOISTURE EXCHANGE FILTERS;
D O I
10.1093/bja/69.5.522
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We have examined the properties of six heat and moisture exchange filters (HMEF) to ascertain their resistance to liquid flow and their ability to retain a challenge bacterium, Pseudomonas diminuta, from aqueous and nebulized suspensions. Only one HMEF, the Pall Ultipor was able to withstand a significantly greater pressure of liquid than that found in clinical practice. However, when breached, the HMEF were unable to prevent transmission of micro-organisms from aqueous suspension. Only the Darex Hydrobac filter failed to meet the manufacturer's claim for filter efficiency for nebulized bacteria, mainly because the filter housing failed under test. When the reduction in bacterial cells after passage of the nebulized Pseudomonas diminuta through the HMEFs was analysed statistically, the data showed that the HMEF produced by Pall (Ultipor) and Intersurgical (Filter therm) were superior to those produced by DAR Mediplan (Hygrobac), Intertech (HME 225-2835-800) and Gibeck (Humid-vent).
引用
收藏
页码:522 / 525
页数:4
相关论文
共 13 条
[1]  
CHALON J, 1984, ANESTH ANALG, V63, P66
[2]  
COHEN L, 1979, STATISTICS ED PHYSIC
[3]  
Duberstein R, 1978, J Parenter Drug Assoc, V32, P192
[4]   CONTAMINATION CONTROL IN LONG-TERM VENTILATION - A CLINICAL-STUDY USING A HEAT-EXCHANGING AND MOISTURE-EXCHANGING FILTER [J].
GALLAGHER, J ;
STRANGEWAYS, JEM ;
ALLTGRAHAM, J .
ANAESTHESIA, 1987, 42 (05) :476-481
[5]   SURGICAL EMPHYSEMA RESULTING FROM A VENTILATOR MALFUNCTION [J].
HILTON, PJ ;
CLEMENT, JA .
ANAESTHESIA, 1983, 38 (04) :342-345
[6]  
Linko K, 1984, Eur J Anaesthesiol, V1, P285
[7]   HEAT AND MOISTURE EXCHANGERS AND VAPORIZING HUMIDIFIERS IN THE INTENSIVE-CARE UNIT [J].
MARTIN, C ;
PERRIN, G ;
GEVAUDAN, MJ ;
SAUX, P ;
GOUIN, F .
CHEST, 1990, 97 (01) :144-149
[8]   COMPLICATIONS RELATED TO THE USE OF A HEAT AND MOISTURE EXCHANGER [J].
PRASAD, KK ;
CHEN, L .
ANESTHESIOLOGY, 1990, 72 (05) :958-958
[9]   A COMPARISON OF 5 HEAT AND MOISTURE EXCHANGERS [J].
SHELLY, M ;
BETHUNE, DW ;
LATIMER, RD .
ANAESTHESIA, 1986, 41 (05) :527-532
[10]  
TATTERSAL M, 1984, ANAESTHESIA POINTS W, V15, P17