The impact of mass density variations on an electron Monte Carlo algorithm for radiotherapy dose calculations

被引:13
作者
Fang, Raymond [1 ]
Mazur, Thomas [1 ]
Mutic, Sasa [1 ]
Khan, Rao [1 ]
机构
[1] Washington Univ, Sch Med, Dept Radiat Oncol, St Louis, MO 63130 USA
关键词
Monte Carlo electron dose calculation; Mass density; Hounsfield units;
D O I
10.1016/j.phro.2018.10.002
中图分类号
R73 [肿瘤学];
学科分类号
100214 [肿瘤学];
摘要
Background and Purpose: A key step in electron Monte Carlo dose calculation requires converting Computed Tomography (CT) numbers from a tomographic acquisition to a mass density. This study investigates the dosimetric consequences of perturbations applied to a calibration table between CT number and mass density. Materials and Methods: A literature search was performed to define lower and upper bounds for physically reasonable perturbations to a reference CT number to mass density calibration table. Electron beam dose was calculated for ten patients using these variations and the results were compared to clinical plans originally derived with a reference calibration table. Dose differences both globally and in the Planning Target Volume (PTV) were assessed using dose- and volume-based metrics and 3-dimensional gamma analysis for each patient. Results: Small but statistically significant differences were observed between perturbations and reference data for certain metrics including volume of the 50% prescription isodose. Upper and lower variations in CT number to mass density calibration yielded mean values of V50% that were 4.4% larger and 2.1% smaller than reference values respectively. Gamma analysis using 3%/3mm criteria indicated > 99% passing rate for the PTV for all patients. Global gamma analysis for some patients showed larger discrepancies possibly due to large electron path lengths through inhomogeneities. Conclusions: In most patients, physically reasonable perturbations in CT number to mass density curves will not induce clinically significant impact on calculated target dose distributions. Strong dependence of electron transport on voxel material may produce dose speckle throughout the volume. Care should be taken in evaluating critical structures at depths beyond the target volume in highly heterogeneous regions.
引用
收藏
页码:1 / 7
页数:7
相关论文
共 37 条
[1]
[Anonymous], 2013, CATPHAN 504 MANUAL
[2]
Dual-energy CT-based material extraction for tissue segmentation in Monte Carlo dose calculations [J].
Bazalova, Magdalena ;
Carrier, Jean-Francois ;
Beaulieu, Luc ;
Verhaegen, Frank .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (09) :2439-2456
[3]
Correction of CT artifacts and its influence on Monte Carlo dose calculations [J].
Bazalova, Magdalena ;
Beaulieu, Luc ;
Palefsky, Steven ;
Verhaegen, Frank .
MEDICAL PHYSICS, 2007, 34 (06) :2119-2132
[4]
Multi-detector row CT attenuation measurements: Assessment of intra- and interscanner variability with an anthropomorphic body CT phantom [J].
Birnbaum, Bernard A. ;
Hindman, Nicole ;
Lee, Julie ;
Babb, James S. .
RADIOLOGY, 2007, 242 (01) :109-119
[5]
Applications of simulator computed tomography number for photon dose calculations during radiotherapy treatment planning [J].
Chu, JCH ;
Ni, B ;
Kriz, R ;
Saxena, VA .
RADIOTHERAPY AND ONCOLOGY, 2000, 55 (01) :65-73
[6]
Dosimetric impact of computed tomography calibration on a commercial treatment planning system for external radiation therapy [J].
Cozzi, L ;
Fogliata, A ;
Buffa, F ;
Bieri, S .
RADIOTHERAPY AND ONCOLOGY, 1998, 48 (03) :335-338
[7]
Scanner and kVp dependence of measured CT numbers in the ACR CT phantom [J].
Cropp, Robert J. ;
Seslija, Petar ;
Tso, David ;
Thakur, Yogesh .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (06) :338-349
[8]
Computed tomography imaging parameters for inhomogeneity correction in radiation treatment planning [J].
Das, Indra J. ;
Cheng, Chee-Wai ;
Cao, Minsong ;
Johnstone, Peter A. S. .
JOURNAL OF MEDICAL PHYSICS, 2016, 41 (01) :3-11
[9]
The indirect use of CT numbers to establish material properties needed for Monte Carlo calculation of dose distributions in patients [J].
du Plessis, FCP ;
Willemse, CA ;
Lotter, MG ;
Goedhals, L .
MEDICAL PHYSICS, 1998, 25 (07) :1195-1201
[10]
CT-ED conversion on a GE Lightspeed-RT scanner: influence of scanner settings [J].
Ebert, M. A. ;
Lambert, J. ;
Greer, P. B. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2008, 31 (02) :154-159