DEVELOPMENT OF A NONLINEAR FINITE-ELEMENT MODELING OF THE BELOW-KNEE PROSTHETIC SOCKET INTERFACE

被引:65
作者
ZHANG, M
LORD, M
TURNERSMITH, AR
ROBERTS, VC
机构
[1] Centre of Rehabilitation Engineering, Department of Medical Engineering and Physics, King's College School of Medicine and Dentistry, London, SE5 9RS, Denmark Hill
关键词
PROSTHESIS; FINITE ELEMENT MODEL; INTERFACE STRESS; SLIP; FRICTION;
D O I
10.1016/1350-4533(95)00002-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A non-linear finite element model has been established to predict the pressure and shear stress distribution at the limb-socket interface in below-knee amputees with consideration of the skin-liner interface friction and slip. In this model, the limb tissue and socket liner were respectively meshed into 954 and 450 three-dimensional eight-node isoparametric brick elements, based on measurements of an individual's amputated limb surface; the bone was meshed into three-dimensional six-node triangular prism elements, based on radiographic measurements of the individual's residual limb. The socket shell was assumed to be a rigid boundary. An important feature of this model is the use of 450 interface elements (ABAQUS INTER4) which mimic the interface friction condition. The results indicate that a maximum pressure of 226 kPa, shear stress of 53 kPa and less than 4 mm slip exist at the skin-liner interface when the full body weight of 800 N is applied to the limb. The results also show that the coefficient of friction is a very sensitive parameter in determining the interface pressures, shear stresses and slip. With the growth of coefficient of friction, the shear stresses will increase, whit the pressure and slip will decrease.
引用
收藏
页码:559 / 566
页数:8
相关论文
共 17 条
[1]  
Appoldt F., 1968, B PROSTHET RES FAL, V3, P106
[2]  
CHILDRESS DL, 1987, J REHABIL RES DEV, V25, P22
[3]   THE SKIN SURFACE AND FRICTION [J].
NAYLOR, PFD .
BRITISH JOURNAL OF DERMATOLOGY, 1955, 67 (07) :239-248
[4]  
Pearson J R, 1973, Bull Prosthet Res, V10, P52
[5]   ANALYSIS OF A BELOW-KNEE PATELLAR TENDON-BEARING PROSTHESIS - A FINITE-ELEMENT STUDY [J].
QUESADA, P ;
SKINNER, HB .
JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT, 1991, 28 (03) :1-12
[6]  
REYNOLDS D, 1988, THESIS U LONDON
[7]   INTERFACE LOAD ANALYSIS FOR COMPUTER-AIDED-DESIGN OF BELOW-KNEE PROSTHETIC SOCKETS [J].
REYNOLDS, DP ;
LORD, M .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1992, 30 (04) :419-426
[8]  
Sanders J. E., 1991, THESIS U WASHINGTON
[9]   INTERFACE SHEAR STRESSES DURING AMBULATION WITH A BELOW-KNEE PROSTHETIC LIMB [J].
SANDERS, JE ;
DALY, CH ;
BURGESS, EM .
JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT, 1992, 29 (04) :1-8
[10]  
SANDERS JE, 1993, J REHABIL RES DEV, V30, P191