FUNCTIONAL RESIDUAL CAPACITY AND VENTILATION HOMOGENEITY IN MECHANICALLY VENTILATED SMALL NEONATES

被引:50
作者
VILSTRUP, CT
BJORKLUND, LJ
LARSSON, A
LACHMANN, B
WERNER, O
机构
[1] UNIV LUND HOSP, DEPT PEDIAT, S-22185 LUND, SWEDEN
[2] ERASMUS UNIV, DEPT ANESTHESIA, 3000 DR ROTTERDAM, NETHERLANDS
关键词
MULTIBREATH WASHOUT; POSITIVE END-EXPIRATORY PRESSURE; GAS DISTRIBUTION;
D O I
10.1152/jappl.1992.73.1.276
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
A modification of a computerized tracer gas (SF6) washout method was designed for serial measurements of functional residual capacity (FRC) and ventilation homogeneity in mechanically ventilated very-low-birth-weight infants with tidal volumes down to 4 ml. The method, which can be used regardless of the inspired O2 concentration, gave accurate and reproducible results in a lung model and good agreement compared with He dilution in rabbits. FRC was measured during 2-4 cmH2O of positive end-expiratory pressure (PEEP) in 15 neonates (700-1,950 g), most of them with mild-to-moderate respiratory distress syndrome. FRC increased with body weight and decreased (P < 0.05) with increasing O2 requirement. Change to zero end-expiratory pressure caused an immediate decrease in FRC by 29% (P < 0.01) and gave FRC (ml) = -1.4 + 17 X weight (kg) (r = 0.83). Five minutes after PEEP was discontinued (n = 12), FRC had decreased by a further 16% (P < 0.01). The washout curves indicated a near-normal ventilation homogeneity not related to changes in PEEP. This was interpreted as evidence against the presence of large volumes of trapped alveolar gas.
引用
收藏
页码:276 / 283
页数:8
相关论文
共 25 条
[1]   PROBLEMS IN MEASUREMENT OF THORACIC GAS VOLUME IN INFANCY [J].
BEARDSMORE, CS ;
STOCKS, J ;
SILVERMAN, M .
JOURNAL OF APPLIED PHYSIOLOGY, 1982, 52 (04) :995-999
[2]   DETERMINATION OF PULMONARY PARENCHYMAL TISSUE VOLUME AND PULMONARY CAPILLARY BLOOD FLOW IN MAN [J].
CANDER, L ;
FORSTER, RE .
JOURNAL OF APPLIED PHYSIOLOGY, 1959, 14 (04) :541-551
[3]   LUNG FUNCTION STUDIES .8. ANALYSIS OF ALVEOLAR VENTILATION BY PULMONARY N-2 CLEARANCE CURVES [J].
FOWLER, WS ;
CORNISH, ER ;
KETY, SS .
JOURNAL OF CLINICAL INVESTIGATION, 1952, 31 (01) :40-50
[4]   EFFECTS OF ENDOTRACHEAL-TUBE LEAKS ON FUNCTIONAL RESIDUAL CAPACITY DETERMINATION IN INTUBATED NEONATES [J].
FOX, WW ;
SCHWARTZ, JG ;
SHAFFER, TH .
PEDIATRIC RESEARCH, 1979, 13 (01) :60-64
[5]   FUNCTIONAL RESIDUAL CAPACITY IN NORMAL NEONATES AND CHILDREN UP TO 5 YEARS OF AGE DETERMINED BY A N-2 WASHOUT METHOD [J].
GERHARDT, T ;
REIFENBERG, L ;
HEHRE, D ;
FELLER, R ;
BANCALARI, E .
PEDIATRIC RESEARCH, 1986, 20 (07) :668-671
[6]   ROLE OF CONDUCTING AIRWAYS IN PARTIAL SEPARATION OF INHALED GAS-MIXTURES [J].
HORSFIELD, K ;
DAVIES, A ;
CUMMING, G .
JOURNAL OF APPLIED PHYSIOLOGY, 1977, 43 (03) :391-396
[7]  
HUGHE SJMB, 1982, B EUR PHYSIOPATHOL R, V18, P309
[8]  
JONES HA, 1982, B EUR PHYSIOPATH RES, V18, P319
[9]  
JONMARKER C, 1986, ANESTHESIOLOGY, V65, P259
[10]   AN ANALYZER FOR IN-LINE MEASUREMENT OF EXPIRATORY SULFUR-HEXAFLUORIDE CONCENTRATION [J].
JONMARKER, C ;
CASTOR, R ;
DREFELDT, B ;
WERNER, O .
ANESTHESIOLOGY, 1985, 63 (01) :84-88