CLINICAL SURVEILLANCE OF AN ACTIVE FIXATION, BIPOLAR, POLYURETHANE INSULATED PACING LEAD .2. THE VENTRICULAR LEAD

被引:12
作者
GLIKSON, M
VONFELDT, LK
SUMAN, VJ
HAYES, DL
机构
[1] MAYO CLIN & MAYO FDN,DIV CARDIOVASC DIS & INTERNAL MED,ROCHESTER,MN 55905
[2] MAYO CLIN & MAYO FDN,BIOSTAT SECT,ROCHESTER,MN 55905
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1994年 / 17卷 / 09期
关键词
PACING LEADS; VENTRICULAR SCREW-IN LEADS; LEAD PERFORMANCE; LEAD SURVIVAL;
D O I
10.1111/j.1540-8159.1994.tb01514.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Since 1989, 72 Telectronics 330-201 active fixation, polyurethane insulated ventricular leads (Accufix) have been implanted at the Mayo Clinic. There were four (5.6%) acute lead related complications (perforation, microdislodgment, and macrodislodgment), three of which led to early reoperation. Over a follow-up time of up to 2.7 years (median 9.4 months), there were six (8.3%) chronic lead related complications but no failures of lead material. Most of these complications developed during the first month, and half of them were transient, with documented improvement later. Two patients (2.8%) required reoperation for chronic complications. At follow-up examination of the pacing thresholds, usually performed about 3 months after implantation, 14.3% of the examined patients had high pacing thresholds necessitating high-output programming. The mechanisms and later evolution of this phenomenon should be further evaluated.
引用
收藏
页码:1499 / 1502
页数:4
相关论文
共 15 条
  • [1] 3-YEAR CLINICAL-EXPERIENCE WITH A NEW ENDOCARDIAL SCREW-IN LEAD WITH INTRODUCTION PROTECTION FOR USE IN THE ATRIUM AND VENTRICLE
    BISPING, HJ
    KREUZER, J
    BIRKENHEIER, H
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1980, 3 (04): : 424 - 435
  • [2] EIGEL P, 1982, CARDIAC PACING ELECT, P1107
  • [3] SURVIVAL OF IMPLANTABLE PACEMAKER LEADS
    FURMAN, S
    BENEDEK, ZM
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1910 - 1914
  • [4] CLINICAL SURVEILLANCE OF AN ACTIVE FIXATION, BIPOLAR, POLYURETHANE INSULATED PACING LEAD .1. THE ATRIAL LEAD
    GLIKSON, M
    VONFELDT, LK
    SUMAN, VJ
    HAYES, DL
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (08): : 1399 - 1404
  • [5] A MULTICENTER EXPERIENCE WITH A BIPOLAR TINED POLYURETHANE VENTRICULAR LEAD
    HAYES, DL
    GRAHAM, KJ
    IRWIN, M
    VIDAILLET, H
    DISLER, G
    SWEESY, M
    OSBORN, MJ
    SUMAN, VJ
    NEUBAUER, SA
    SEEBANDT, M
    KALLINEN, L
    CROWSON, CS
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (07): : 1033 - 1039
  • [6] COMPLICATIONS OF PERMANENT TRANSVENOUS CARDIAC PACING - A 14-YEAR REVIEW OF ALL TRANSVENOUS PACEMAKERS INSERTED AT ONE COMMUNITY-HOSPITAL
    HILL, PE
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (03): : 564 - 570
  • [7] HOLMES DR, 1980, J THORAC CARDIOV SUR, V79, P565
  • [8] KAYE G, 1990, EUR J CARDIAC PACING, V1, P154
  • [9] PERRINS EJ, 1981, BRIT HEART J, V46, P196
  • [10] TIME COURSE OF TRANSVENOUS PACEMAKER STIMULATION IMPEDANCE, CAPTURE THRESHOLD, AND ELECTROGRAM AMPLITUDE
    PLATIA, EV
    BRINKER, JA
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (05): : 620 - 625