VERIFICATION OF GAMMA-SPECTROSCOPY PROGRAMS - A STANDARDIZED APPROACH

被引:10
作者
KOSKELO, MJ [1 ]
MERCIER, MT [1 ]
机构
[1] CANBERRA NUCL PROD GRP,SCHAUMBURG,IL 60196
关键词
D O I
10.1016/0168-9002(90)90799-C
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this paper, the basic concepts contained in a proposed revision of the ANSI 42.14 standard for testing gamma spectroscopy software are used to evaluate peak search and area calculation algorithms currently in use in Canberra and Nuclear Data software packages, and show that they indeed meet the needs of most users. A discussion on the limitations of the algorithms within the scope of the tests is also provided.
引用
收藏
页码:318 / 321
页数:4
相关论文
共 7 条
[1]   QUALITY ASSURANCE IN THE DETERMINATION OF OVERLAPPING PEAK AREAS [J].
CHRISTENSEN, LH ;
HEYDORN, K .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1987, 113 (01) :19-34
[2]   VERIFICATION TESTING OF COMMERCIALLY AVAILABLE COMPUTER-PROGRAMS FOR PHOTOPEAK AREA EVALUATION [J].
HEYDORN, K ;
CHRISTENSEN, LH .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1988, 124 (02) :467-480
[3]  
HOUTERMANS H, 1990, IAEA G2 REP
[4]  
PARR RM, 1979, US DEP ENERGY S SERI, V49, P544
[5]   PHOTOPEAK METHOD FOR COMPUTER ANALYSIS OF GAMMA-RAY SPECTRA FROM SEMICONDUCTOR DETECTORS [J].
ROUTTI, JT ;
PRUSSIN, SG .
NUCLEAR INSTRUMENTS & METHODS, 1969, 72 (02) :125-&
[6]  
SANDERSON CG, 1987, EVALUATION COMMERCIA
[7]  
1990, ANSI4214 STAND