MOSSBAUER LINESHAPE DISTORTIONS DUE TO FINITE DIMENSIONS OF SOURCE AND DETECTOR

被引:1
作者
FLORESLLAMAS, H [1 ]
ZAMORANOULLOA, R [1 ]
机构
[1] ESCUELA SUPER FIS & MATEMAT,INST POLITECN NACL,DEPT FIS,MEXICO CITY,DF,MEXICO
关键词
D O I
10.1016/0168-583X(91)95600-I
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The effects of non-collimation of gamma-rays and finite dimensions of source and detector upon the isomer shift, line broadening and line height of Mossbauer spectra have been determined by means of a two-dimensional angular distribution f(r, rho) that weights a Lorentzian lineshape function. This distribution function, along with a few approximations, allow one to calculate, with ease, these Mossbauer spectral parameters. Our expressions, valid for finite dimensions, of source, detector and absorber, generalize previous calculations. As a result, we can say that when a finite source (R(s) = 0.4 cm) is 10 cm distance from a detector (R(d) = 1.4 cm), the non-collimation of the gamma-rays generates distortions of the absorption line that are still within the experimental error.
引用
收藏
页码:272 / 277
页数:6
相关论文
共 13 条
[11]   VELOCITY SHIFTS IN MOSSBAUER SPECTROSCOPY [J].
HERSHKOWITZ, N .
NUCLEAR INSTRUMENTS & METHODS, 1967, 53 (01) :172-+
[12]  
RIESH A, 1986, NUCL INSTRUM METH A, V245, P162
[13]  
SPIJKERMAN JJ, 1965, MOSSBAUER EFFECT MET, V1, P115, DOI 10.1007/978-1-4757-1541-5_10