THERMAL-DIFFUSION PROBE AND INSTRUMENT SYSTEM FOR TISSUE BLOOD-FLOW MEASUREMENTS - VALIDATION IN PHANTOMS AND IN-VIVO ORGANS

被引:15
作者
DELHOMME, G
NEWMAN, WH
ROUSSEL, B
JOUVET, M
BOWMAN, HF
DITTMAR, A
机构
[1] MIT, HARVARD MIT DIV HLTH SCI & TECHNOL, CAMBRIDGE, MA 02139 USA
[2] THERMAL TECHNOL INC, CAMBRIDGE, MA 02142 USA
[3] UNIV LYON 1, DEPT EXPTL MED, CNRS, UA 1195, INSERM, U52, F-69373 LYON 08, FRANCE
关键词
D O I
10.1109/10.301732
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A minimally invasive probe and instrument system for real-time measurements of temperature, thermal conductivity and tissue blood flow has been designed for research and clinical use. The essence of the probe is a thermistor, located at the tip of catheters or glass and steel needles, and operating in transient self-heated mode at constant temperature increment. Thermal conductivity and tissue blood how are determined by use of a coupled tissue-probe thermal model. The effects of temporal baseline temperature shifts are minimized by a novel, automatic, analog compensation circuit. Very short healing periods (3 s) and cooling periods (12 s) provided near-continuous measurements (4/min). Calibration experiments performed in media of known thermal conductivity exhibit a linear response with respect to thermal conductivity. In vitro experiments performed in isolated perfused dog liver preparations are presented to evaluate this instrument system. In vivo experiments performed in cat brain, dog liver, and human tumor demonstrate the ability of this instrument system to perform physiologically valid measurements (comparison inter-subjects and intra-subjeets). The minimally invasive probes (0.8 mm OD) are capable of long term measurements (several months), with minimal tissue reactions (0.3 mm around the probe).
引用
收藏
页码:656 / 662
页数:7
相关论文
共 19 条
[11]  
Gibbs FA, 1933, P SOC EXP BIOL MED, V31, P141
[12]   INTERNAL CALORIMETRY IN THE DETERMINATION OF THERMAL CONDUCTIVITY AND BLOOD FLOW [J].
GRAYSON, J .
JOURNAL OF PHYSIOLOGY-LONDON, 1952, 118 (01) :54-72
[13]   THERMAL-CONDUCTIVITY AND H2O CONTENT IN RABBIT KIDNEY CORTEX AND MEDULLA [J].
HOLMES, KR ;
RYAN, W ;
CHEN, MM .
JOURNAL OF THERMAL BIOLOGY, 1983, 8 (04) :311-313
[14]   THERMAL-PROPERTIES OF TISSUE EQUIVALENT PHANTOM MATERIALS [J].
LEONARD, JB ;
FOSTER, KR ;
ATHEY, TW .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1984, 31 (07) :533-536
[15]  
NEWMAN WH, 1986, 8TH P ANN IEEE C ENG, V3, P1503
[17]   THERMAL CONDUCTIVITY MEASUREMENTS AND PREDICTIONS FOR BIOLOGICAL FLUIDS AND TISSUES [J].
POPPENDIEK, HF ;
RANDALL, R ;
BREEDEN, JA ;
CHAMBERS, JE ;
MURPHY, JR .
CRYOBIOLOGY, 1967, 3 (04) :318-+
[18]  
Roussel B, 1980, Waking Sleeping, V4, P63
[19]   THE SIMULTANEOUS MEASUREMENT OF THERMAL-CONDUCTIVITY, THERMAL-DIFFUSIVITY, AND PERFUSION IN SMALL VOLUMES OF TISSUE [J].
VALVANO, JW ;
ALLEN, JT ;
BOWMAN, HF .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1984, 106 (03) :192-197