A COMPARISON OF LYOPHILIZED ARTIFICIALLY DEPLETED PLASMAS AND LYOPHILIZED PLASMAS FROM PATIENTS RECEIVING WARFARIN IN CORRECTING FOR COAGULOMETER EFFECTS ON INTERNATIONAL NORMALIZED RATIOS

被引:33
作者
POLLER, L
TRIPLETT, DA
HIRSH, J
CARROLL, J
CLARKE, K
机构
[1] BALL MEM HOSP,MUNCIE,IN 47303
[2] HAMILTON CIV HOSP,RES CTR,HAMILTON,ON,CANADA
[3] ORTHO DIAGNOST SYST INC,RARITAN,NJ
关键词
WARFARIN; COUMADIN; THROMBOPLASTINS; PROTHROMBIN TIME; LABORATORY STANDARDIZATION;
D O I
10.1093/ajcp/103.3.366
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
The comparative value of lyophilized artificially depleted plasmas and plasmas from warfarin-treated patients in local (thromboplastin/coagulometer) system International Sensitivity Indices (ISI) determination has been studied at 39 centers in an international collaborative study. At each center, both types of plasmas were tested with two common thromboplastins (low ISI RecombiPlasTin and high ISI OBT), and in-house reagents using three brands of coagulometers. In 95 local systems, they was agreement. With both thromboplastins, coagulometers lowered the ISI from the manual or stated ISI with all systems. Differences between the two types of lyophilized plasmas with the two common thromboplastins were not of clinical importance. Artificially depleted plasmas are easier to obtain in sufficient volumes than plasma from, warfarin-treated patients, nod more easily provide the wide range of INR values required for reliable local ISI calibration. The agreement between the two types of lyophilized calibrant is reassuring.
引用
收藏
页码:366 / 371
页数:6
相关论文
共 13 条
[1]   ASSESSMENT OF VALUE OF CALIBRATED LYOPHILIZED PLASMAS TO DETERMINE INTERNATIONAL SENSITIVITY INDEX FOR COAGULOMETERS [J].
CLARKE, K ;
TABERNER, DA ;
THOMSON, JM ;
MORRIS, JA ;
POLLER, L .
JOURNAL OF CLINICAL PATHOLOGY, 1992, 45 (01) :58-60
[2]   COMPARISON OF 2 AUTOMATED COAGULOMETERS AND THE MANUAL TILT-TUBE METHOD FOR THE DETERMINATION OF PROTHROMBIN TIME [J].
DANGELO, A ;
SEVESO, MP ;
DANGELO, SV ;
GILARDONI, F ;
MACAGNI, A ;
MANOTTI, C ;
BONINI, P .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1989, 92 (03) :321-328
[3]  
HOUBOUYAN LL, 1993, THROMB HAEMOSTASIS, V69, P663
[4]  
POGGIO M, 1988, THROMB HAEMOSTASIS, V62, P868
[5]   EFFECT OF THE CHOICE OF WHO INTERNATIONAL REFERENCE PREPARATION FOR THROMBOPLASTIN ON INTERNATIONAL NORMALIZED RATIOS [J].
POLLER, L ;
TABERNER, DA ;
THOMSON, JM ;
MORRIS, J ;
MIBASHAN, RS ;
SHINTON, NK .
JOURNAL OF CLINICAL PATHOLOGY, 1993, 46 (01) :64-66
[6]   EFFECT OF AUTOMATION ON PROTHROMBIN TIME TEST IN NEQAS SURVEYS [J].
POLLER, L ;
THOMSON, JM ;
TABERNER, DA .
JOURNAL OF CLINICAL PATHOLOGY, 1989, 42 (01) :97-100
[7]   THE CORRECTION OF COAGULOMETER EFFECTS ON INTERNATIONAL NORMALIZED RATIOS - A MULTICENTER EVALUATION [J].
POLLER, L ;
THOMSON, JM ;
TABERNER, DA ;
CLARKE, DK .
BRITISH JOURNAL OF HAEMATOLOGY, 1994, 86 (01) :112-117
[8]  
POLLER L, 1995, AM J CLIN PATHOL, V103
[9]  
RAY MJ, 1990, THROMB HAEMOSTASIS, V63, P424
[10]  
Schmidt Nico A, 1989, CLIN CHEM, V355, P840