T-1 MEASUREMENTS WITH CLINICAL MR UNITS

被引:9
作者
TAMURA, H [1 ]
YANAGAWA, I [1 ]
HIKICHI, T [1 ]
MATSUMOTO, K [1 ]
TAKAHASHI, S [1 ]
SAKAMOTO, K [1 ]
机构
[1] TOHOKU ROSAI HOSP, DEPT RADIOL, SENDAI, MIYAGI 981, JAPAN
关键词
MAGNETIC RESONANCE IMAGING; PROTON SPIN-LATTICE RELAXATION TIME MEASUREMENT;
D O I
10.1620/tjem.175.249
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To improve accuracy and feasibility of the proton spin-lattice relaxation time (T-1) measurements in routine Magnetic Resonance Imaging (MRI), a phantom with polyvinyl alcohol gels and gadrinium diethylene triamine pentaacetic acid solutions was imaged with 1.5 T clinical MR units and T(1)s were calculated by fitting several models to the observed signal intensities. When an inversion-recovery sequence was applied, we were able to measure T-1 with a high degree of precision by taking into account the imperfection of the inversion pulse. With a Turbo-fast low angle shot sequence, we were able to obtain T-1 in a short time, but low signal-to-noise ratio limited the precision. Signed intensity with XE sequence was susceptible to variation with the effective flip angle of 90 degrees RF pulse and not suited for T-1 measurement.
引用
收藏
页码:249 / 267
页数:19
相关论文
共 15 条
[1]  
BABU M, 1993, PHYSIOL CHEM PHYS M, V25, P113
[2]  
BLOCH F, 1946, PHYS REV, V70, P460, DOI 10.1103/PhysRev.70.460
[3]   SPIN-LATTICE RELAXATION-TIME MEASUREMENT BY MEANS OF A TURBOFLASH TECHNIQUE [J].
BLUML, S ;
SCHAD, LR ;
STEPANOW, B ;
LORENZ, WJ .
MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (03) :289-295
[4]   QUANTITATIVE MR-IMAGING OF LUMBAR INTERVERTEBRAL DISKS AND VERTEBRAL BODIES - INFLUENCE OF DIURNAL WATER-CONTENT VARIATIONS [J].
BOOS, N ;
WALLIN, A ;
GBEDEGBEGNON, T ;
AEBI, M ;
BOESCH, C .
RADIOLOGY, 1993, 188 (02) :351-354
[5]   A REVIEW OF H-1 NUCLEAR-MAGNETIC-RESONANCE RELAXATION IN PATHOLOGY - ARE T1 AND T2 DIAGNOSTIC [J].
BOTTOMLEY, PA ;
HARDY, CJ ;
ARGERSINGER, RE ;
ALLENMOORE, G .
MEDICAL PHYSICS, 1987, 14 (01) :1-37
[6]  
BOTTOMLEY PA, 1984, MED PHYS, V11, P425, DOI 10.1118/1.595535
[7]   THE USE OF AN IMPROVED INVERSION PULSE WITH THE SPIN-ECHO INVERSION-RECOVERY SEQUENCE TO GIVE INCREASED ACCURACY AND REDUCED IMAGING TIME FOR T1 MEASUREMENTS [J].
GOWLAND, PA ;
LEACH, MO ;
SHARP, JC .
MAGNETIC RESONANCE IN MEDICINE, 1989, 12 (02) :261-267
[8]   SNAPSHOT FLASH MRI - APPLICATIONS TO T1, T2, AND CHEMICAL-SHIFT IMAGING [J].
HAASE, A .
MAGNETIC RESONANCE IN MEDICINE, 1990, 13 (01) :77-89
[9]   THE IMPORTANCE OF PHASE-ENCODING ORDER IN ULTRA-SHORT TR SNAPSHOT MR IMAGING [J].
HOLSINGER, AE ;
RIEDERER, SJ .
MAGNETIC RESONANCE IN MEDICINE, 1990, 16 (03) :481-488
[10]  
RIEDERER SJ, 1990, TISSUE CHARACTERIZAT, P14