MICROFABRICATED SENSOR ARRAYS SENSITIVE TO PH AND K+ FOR IONIC DISTRIBUTION MEASUREMENTS IN THE BEATING HEART

被引:134
作者
COSOFRET, VV
ERDOSY, M
JOHNSON, TA
BUCK, RP
ASH, RB
NEUMAN, MR
机构
[1] UNIV N CAROLINA,DEPT CHEM,CHAPEL HILL,NC 27599
[2] UNIV N CAROLINA,DEPT MED,CHAPEL HILL,NC 27599
[3] N CAROLINA STATE UNIV,DEPT ELECT & COMP ENGN,RALEIGH,NC 27695
[4] CASE WESTERN RESERVE UNIV,METROHLTH MED CTR,DEPT OB GYN,CLEVELAND,OH 44109
关键词
D O I
10.1021/ac00106a001
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Thick-film H+- and K+-selective sensor arrays based on solvent polymeric neutral carrier membranes are described, The sensor away design and construction allow inexpensive and mass production of disposable miniaturized probes useful in the biomedical held, especially for cardiovascular research and monitoring. They are also suitable for use as multisensor electrochemical detectors within modern now injection analysis instrumentation. The carriers used for membrane preparation are H+-selective ETH 5294 chromoionophore and the well-known K+-selective ionophore, valinomycin. The preliminary in vivo experiments, to monitor H+ and K+ ionic distribution in the beating pig heart, help to determine the unique ionic conduction responsible for ventricullar fibrillation in the setting of acute regional myocardial ischemia.
引用
收藏
页码:1647 / 1653
页数:7
相关论文
共 40 条
  • [31] LINDNER E, 1994, DIAGNOSTIC BIOSENSOR, P149
  • [32] LIU CC, 1992, SELECT ELECTR REV, V14, P147
  • [33] MCKINLEY BA, 1984, ION SEL ELECTRODE R, V6, P173
  • [34] MEIER PC, 1980, MED BIOL APPLICATION, P13
  • [35] LIFE TIME OF NEUTRAL CARRIER BASED ION-SELECTIVE LIQUID-MEMBRANE ELECTRODES
    OESCH, U
    SIMON, W
    [J]. ANALYTICAL CHEMISTRY, 1980, 52 (04) : 692 - 700
  • [36] MECHANISMS UNDERLYING THE DEVELOPMENT OF VENTRICULAR-FIBRILLATION DURING EARLY MYOCARDIAL-ISCHEMIA
    POGWIZD, SM
    CORR, PB
    [J]. CIRCULATION RESEARCH, 1990, 66 (03) : 672 - 695
  • [37] Pungor E., 1979, PURE APPL CHEM, V51, P1913
  • [38] THE SUBENDOCARDIAL BORDER ZONE DURING ACUTE-ISCHEMIA OF THE RABBIT HEART - AN ELECTROPHYSIOLOGIC, METABOLIC, AND MORPHOLOGICAL CORRELATIVE STUDY
    WILENSKY, RL
    TRANUMJENSEN, J
    CORONEL, R
    WILDE, AAM
    FIOLET, JWT
    JANSE, MJ
    [J]. CIRCULATION, 1986, 74 (05) : 1137 - 1146
  • [39] DEVELOPMENT OF CHEMICAL SENSORS USING MICROFABRICATION AND MICROMACHINING TECHNIQUES
    WU, QH
    LEE, KM
    LIU, CC
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1993, 13 (1-3) : 1 - 6
  • [40] POLYMER MEMBRANE-BASED ION-SELECTIVE, GAS-SELECTIVE AND BIO-SELECTIVE POTENTIOMETRIC SENSORS
    YIM, HS
    KIBBEY, CE
    MA, SC
    KLIZA, DM
    LIU, D
    PARK, SB
    TORRE, CE
    MEYERHOFF, ME
    [J]. BIOSENSORS & BIOELECTRONICS, 1993, 8 (01) : 1 - 38