20-ELEMENT HGI2 ENERGY DISPERSIVE-X-RAY ARRAY DETECTOR SYSTEM

被引:7
作者
IWANCZYK, JS
DORRI, N
WANG, M
SZCZEBIOT, RW
DABROWSKI, AJ
HEDMAN, B
HODGSON, KO
PATT, BE
机构
[1] SLAC,STANFORD SYNCHROTRON RADIAT LAB,STANFORD,CA 94309
[2] STANFORD UNIV,DEPT CHEM,STANFORD,CA 94305
[3] EG&G ENERGY MEASUREMENTS INC,GOLETA,CA 93117
基金
美国国家卫生研究院; 美国国家航空航天局;
关键词
D O I
10.1109/23.173190
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper describes recent progress in the development of HgI2 energy dispersive x-ray detector arrays and associated miniaturized processing electronics for synchrotron radiations research applications. The experimental results with a 20-element array detector were obtained under realistic synchrotron beam conditions at SSRL. An energy resolution of 250 eV (FWHM) at 5.9 keV (Mn-Kalpha) was achieved. Energy resolution and throughput measurements versus input count rate and energy of incoming radiation have been measured. Extended X-ray Absorption Fine Structure (EXAFS) spectra were taken from diluted samples simulating proteins with nickel.
引用
收藏
页码:1275 / 1280
页数:6
相关论文
共 10 条
[1]   A 13-ELEMENT GE DETECTOR FOR FLUORESCENCE EXAFS [J].
CRAMER, SP ;
TENCH, O ;
YOCUM, M ;
GEORGE, GN .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 266 (1-3) :586-591
[2]  
CRAMER SP, 1991, XRAY ABSORPTION FINE, P640
[3]   ADVANCES IN THE DEVELOPMENT OF ENCAPSULANTS FOR MERCURIC IODIDE X-RAY-DETECTORS [J].
IWANCZYK, JS ;
WANG, YJ ;
BRADLEY, JG ;
ALBEE, AL ;
SCHNEPPLE, WF .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (06) :2214-2218
[4]   ADVANCES IN MERCURIC IODIDE ENERGY DISPERSIVE-X-RAY ARRAY DETECTORS AND ASSOCIATED MINIATURIZED PROCESSING ELECTRONICS [J].
IWANCZYK, JS ;
DORRI, N ;
WANG, M ;
SZAWLOWSKI, M ;
WARBURTON, WK ;
HEDMAN, B ;
HODGSON, KO .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (07) :1561-1567
[5]   THE HGI2 ENERGY DISPERSIVE-X-RAY ARRAY DETECTORS AND MINIATURIZED PROCESSING ELECTRONICS PROJECT [J].
IWANCZYK, JS ;
DORRI, N ;
WANG, M ;
SZAWLOWSKI, M ;
PATT, BE ;
WARBURTON, WK ;
HEDMAN, B ;
HODGSON, KO .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :198-202
[6]   PERFORMANCE AND DURABILITY OF HGI2 X-RAY-DETECTORS FOR SPACE MISSIONS [J].
IWANCZYK, JS ;
WANG, YJ ;
BRADLEY, JG ;
CONLEY, JM ;
ALBEE, AL ;
ECONOMOU, TE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) :841-845
[7]   AN ULTRA LOW-NOISE, LOW-POWER HYBRID PREAMPLIFIER FOR X-RAY SPECTROMETRY [J].
IWANCZYK, JS ;
DABROWSKI, AJ ;
DANCY, BW ;
PATT, BE ;
DEVORE, TM ;
DELDUCA, A ;
ORTALE, C ;
SCHNEPPLE, WF ;
BARKSDALE, JE ;
THOMPSON, TJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1987, 34 (01) :124-126
[8]   CAPABILITIES OF MERCURIC IODIDE AS A ROOM-TEMPERATURE X-RAY DETECTOR [J].
SLAPA, M ;
HUTH, GC ;
SEIBT, W ;
SCHIEBER, MM ;
RANDTKE, PT .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1976, 23 (01) :102-111
[9]   THEORY OF EXTENDED X-RAY-ABSORPTION FINE-STRUCTURE [J].
STERN, EA .
PHYSICAL REVIEW B, 1974, 10 (08) :3027-3037
[10]   RECENT ADVANCES WITH HGI2 X-RAY DETECTORS [J].
SWIERKOW.SP ;
ARMANTRO.GA ;
WICHNER, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1974, NS21 (01) :302-304