3-DIMENSIONAL TIME-OF-FLIGHT MR ANGIOGRAPHY - APPLICATIONS IN THE ABDOMEN AND THORAX
被引:83
作者:
LEWIN, JS
[1
]
LAUB, G
[1
]
HAUSMANN, R
[1
]
机构:
[1] SIEMENS AG,MED ENGN GRP,W-8520 ERLANGEN,GERMANY
来源:
关键词:
MAGNETIC RESONANCE;
ANGIOGRAPHY;
ARTERIES;
MR STUDIES;
MAGNETIC RESONANCE (MR);
PULSE SEQUENCES;
RAPID IMAGING;
TECHNOLOGY;
D O I:
10.1148/radiology.179.1.2006288
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
The application of three-dimensional (3D) time-of-flight (TOF) magnetic resonance (MR) angiographic techniques to the vasculature of the abdomen and thorax has not, to the authors' knowledge, been previously reported; this is possibly due to the large amount of physiologic motion in these regions along with the anticipated sensitivity of the 3D acquisition scheme to image degradation caused by motion artifact. The authors describe an asymmetric short-echo velocity-compensated 3D TOF technique that minimized the effects of physiologic motion on image quality and provided rapid high-resolution 3D MR arteriograms of the abdomen and thorax. Contiguous 3D volumes were often combined to provide sufficient anatomic coverage. Benefits include high spatial resolution and minimization of signal loss; limitations include sensitivity to motion artifact and progressive spin saturation.
引用
收藏
页码:261 / 264
页数:4