NEUTRON-CAPTURE PROMPT GAMMA-RAY ACTIVATION-ANALYSIS AT THE NIST COLD NEUTRON RESEARCH FACILITY

被引:36
作者
LINDSTROM, RM
ZEISLER, R
VINCENT, DH
GREENBERG, RR
STONE, CA
MACKEY, EA
ANDERSON, DL
CLARK, DD
机构
[1] US FDA,WASHINGTON,DC 20204
[2] CORNELL UNIV,ITHACA,NY 14853
来源
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES | 1993年 / 167卷 / 01期
关键词
D O I
10.1007/BF02035470
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An instrument for neutron capture prompt gamma-ray activation analysis (PGAA) has been constructed as part of the Cold Neutron Research Facility at the 20 MW National Institute of Standards and Technology Research Reactor. The neutron fluence rate (thermal equivalent) is 1.5 . 10(8) n . cm-2 . s-1, with negligible fast neutrons and gamma-rays. With compact geometry and hydrogen-free construction, the sensitivity is sevenfold better than an existing thermal instrument. Hydrogen background is thirtyfold lower.
引用
收藏
页码:121 / 126
页数:6
相关论文
共 17 条
[1]   FACILITY FOR NONDESTRUCTIVE ANALYSIS FOR MAJOR AND TRACE-ELEMENTS USING NEUTRON-CAPTURE GAMMA-RAY SPECTROMETRY [J].
ANDERSON, DL ;
FAILEY, MP ;
ZOLLER, WH ;
WALTERS, WB ;
GORDON, GE ;
LINDSTROM, RM .
JOURNAL OF RADIOANALYTICAL CHEMISTRY, 1981, 63 (01) :97-119
[2]  
ANDERSON DL, 1990, BIOL TRACE ELEM RES, V27, P613
[3]   NEUTRON-SCATTERING AND ITS EFFECT ON REACTION-RATES IN NEUTRON ABSORPTION EXPERIMENTS [J].
COPLEY, JRD ;
STONE, CA .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 281 (03) :593-604
[4]   NEUTRON-CAPTURE PROMPT GAMMA-RAY ACTIVATION-ANALYSIS FOR MULTIELEMENT DETERMINATION IN COMPLEX SAMPLES [J].
FAILEY, MP ;
ANDERSON, DL ;
ZOLLER, WH ;
GORDON, GE ;
LINDSTROM, RM .
ANALYTICAL CHEMISTRY, 1979, 51 (13) :2209-2221
[5]  
KITTO ME, 1987, ACS SYM SER, V349, P84
[6]  
LAMAZUE GP, 1991, UNPUB J RADIOANAL NU
[7]   ACTIVATION-ANALYSIS OPPORTUNITIES USING COLD NEUTRON BEAMS [J].
LINDSTROM, RM ;
ZEISLER, R ;
ROSSBACH, M .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1987, 112 (02) :321-330
[8]   MEV NEUTRON-PRODUCTION FROM THERMAL-NEUTRON CAPTURE IN LI-COMPOUNDS AND B-COMPOUNDS [J].
LONE, MA ;
SANTRY, DC ;
INGLIS, WM .
NUCLEAR INSTRUMENTS & METHODS, 1980, 174 (03) :521-529
[9]   EFFECTS OF TARGET SHAPE AND NEUTRON-SCATTERING ON ELEMENT SENSITIVITIES FOR NEUTRON-CAPTURE PROMPT GAMMA-RAY ACTIVATION-ANALYSIS [J].
MACKEY, EA ;
GORDON, GE ;
LINDSTROM, RM ;
ANDERSON, DL .
ANALYTICAL CHEMISTRY, 1991, 63 (03) :288-292
[10]   THE NEUTRON MICROGUIDE AS A PROBE FOR MATERIALS ANALYSIS [J].
MILDNER, DFR .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 299 (1-3) :416-419