REACTION-RATE MEASUREMENTS BY USE OF MEMBRANELESS FLOW-THROUGH BIOSENSORS

被引:4
作者
FERNANDEZROMERO, JM
DECASTRO, MDL
VALCARCEL, M
机构
[1] Department of Analytical Chemistry, Faculty of Sciences, University of Córdoba
关键词
BIOSENSORS; FLOW INJECTION ANALYSIS; IMMOBILIZED ENZYMES; REACTION-RATE MEASUREMENTS;
D O I
10.1016/0925-4005(93)87007-C
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Three different approaches to performing reaction-rate measurements by using flow-through optical biosensors are proposed. This type of biosensor consists of a special flow-cell located in a conventional spectrophotometer, on the inner walls of which a biocatalyst is immobilized. Reaction-rate measurements can be made by coupling the biosensor on-line to different flow-injection (FI) manifolds: a normal FI system in which the flow is halted while the sample plug is in the flow-cell; another where the flow direction is changed iteratively; and an open-closed configuration. The performance of the three manifolds is critically studied and compared by using the same biosensor (immobilized beta-D-glucuronidase) for the determination of glucuronide derivatives.
引用
收藏
页码:203 / 208
页数:6
相关论文
共 14 条
[1]  
BERGMEYER HU, 1983, METHOD ENZYMAT AN, P246
[2]   IMMOBILIZED BETA-GLUCURONIDASE AS AN ONLINE PRECOLUMN MODIFICATION REAGENT FOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
BOWERS, LD ;
JOHNSON, PR .
ANALYTICAL BIOCHEMISTRY, 1981, 116 (01) :111-115
[3]   INTEGRATION OF SEPARATION AND DETECTION IN CONTINUOUS-FLOW SYSTEMS [J].
DECASTRO, MDL ;
VALCARCEL, M .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1991, 10 (04) :114-121
[4]  
DECASTRO MDL, 1989, TRAC-TREND ANAL CHEM, V8, P172
[5]  
DECASTRO MDL, IN PRESS BIOCH SENSO
[6]  
FERNANDEZROMERO JM, IN PRESS ANAL CHEM
[7]   DETERMINATION OF CONJUGATED GLUCURONIC-ACID BY COMBINING ENZYMATIC-HYDROLYSIS WITH LUCIGENIN CHEMI-LUMINESCENCE [J].
KLOPF, LL ;
NIEMAN, TA .
ANALYTICAL CHEMISTRY, 1985, 57 (01) :46-51
[8]   INTEGRATED REACTION AND PHOTOMETRIC DETECTION WITH ENZYMES IMMOBILIZED IN THE FLOW CELL OF A FLOW-INJECTION SYSTEM [J].
LINARES, P ;
DECASTRO, L ;
VALCARCEL, M .
ANALYTICA CHIMICA ACTA, 1990, 230 (01) :199-202
[9]   DETERMINATION OF CHOLESTEROL BY FLOW-INJECTION ANALYSIS WITH IMMOBILIZED CHOLESTEROL OXIDASE [J].
MASOOM, M ;
TOWNSHEND, A .
ANALYTICA CHIMICA ACTA, 1985, 174 (AUG) :293-297
[10]  
MASOOM M, 1984, ANAL CHIM ACTA, V166, P111