COMPARISON OF OSCILLOMETRIC AND INTRAARTERIAL BLOOD-PRESSURE AND PULSE MEASUREMENT

被引:11
作者
RITHALIA, SVS [1 ]
EDWARDS, D [1 ]
机构
[1] MANCHESTER ROYAL INFIRM,INTENS THERAPY UNIT,MANCHESTER M13 9WL,LANCS,ENGLAND
关键词
D O I
10.3109/03091909409030249
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Non-invasive oscillometric blood pressure and pulse measured by an Omron HEM-703CP monitor were compared with arterial values obtained from direct measurements of the radial artery. An excellent correlation and agreement was found between the two methods (systolic r = 0.99; diastolic r = 0.97; pulse r = 0.99), although (here was some variability among individual subjects. The range of difference between them was 0 to 10 mmHg for systolic and - 6 to + 5 mmHg for diastolic pressures. When tested on the bench wing the Metron QA-1280 non-invasive blood pressure analyser the HEM-703CP monitor rarely exhibited errors exceeding 2-3 mmHg over a measurement range of 50-200 mmHg.
引用
收藏
页码:179 / 181
页数:3
相关论文
共 22 条
  • [1] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [2] SPHYGMOMANOMETERS IN HOSPITAL AND FAMILY-PRACTICE - PROBLEMS AND RECOMMENDATIONS
    BURKE, MJ
    TOWERS, HM
    OMALLEY, K
    FITZGERALD, DJ
    OBRIEN, ET
    [J]. BRITISH MEDICAL JOURNAL, 1982, 285 (6340) : 469 - 471
  • [3] DEFECTS OF SPHYGMOMANOMETERS - IMPORTANT SOURCE OF ERROR IN BLOOD-PRESSURE RECORDING
    CONCEICAO, S
    WARD, MK
    KERR, DNS
    [J]. BRITISH MEDICAL JOURNAL, 1976, 1 (6014) : 886 - 888
  • [4] MEASUREMENT ERROR IN THE HAWKSLEY RANDOM ZERO SPHYGMOMANOMETER - WHAT DAMAGE HAS BEEN DONE AND WHAT CAN WE LEARN
    CONROY, RM
    OBRIEN, E
    OMALLEY, K
    ATKINS, N
    [J]. BRITISH MEDICAL JOURNAL, 1993, 306 (6888) : 1319 - 1322
  • [5] CUNNINGHAM S, 1992, BRIT J INTENSIVE CAR, V2, P381
  • [6] DIRECT BLOOD-PRESSURE MEASUREMENT - DYNAMIC-RESPONSE REQUIREMENTS
    GARDNER, RM
    [J]. ANESTHESIOLOGY, 1981, 54 (03) : 227 - 236
  • [7] THE ACCURACY OF FINAPRES NONINVASIVE MEAN ARTERIAL-PRESSURE MEASUREMENTS IN ANESTHETIZED PATIENTS
    GIBBS, NM
    LARACH, DR
    DERR, JA
    [J]. ANESTHESIOLOGY, 1991, 74 (04) : 647 - 652
  • [8] ASSESSMENT OF THE ACCURACY AND ROLE OF SELF-RECORDED BLOOD PRESSURES IN THE MANAGEMENT OF HYPERTENSION
    GOULD, BA
    KIESO, HA
    HORNUNG, R
    ALTMAN, DG
    CASHMAN, PMM
    RAFTERY, EB
    [J]. BRITISH MEDICAL JOURNAL, 1982, 285 (6356) : 1691 - 1694
  • [9] CLINICAL-EVALUATION OF THE PHILIPS-5306B ELECTRONIC BLOOD-PRESSURE MONITOR
    HALL, CL
    GOODFELLOW, J
    WAITES, J
    [J]. JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1991, 15 (4-5) : 177 - 179
  • [10] HALL GE, 1979, EDUC HORIZONS, V4, P202