HIGH-SENSITIVITY NMR MEASUREMENTS AT LOW-TEMPERATURE AND FREQUENCY

被引:11
作者
RUUTU, V
KOIVUNIEMI, J
PARTS, U
HIRAI, A
KRUSIUS, M
机构
[1] Low Temperature Laboratory, Helsinki University of Technology
来源
PHYSICA B | 1994年 / 194卷 / pt 1期
基金
芬兰科学院;
关键词
D O I
10.1016/0921-4526(94)90409-X
中图分类号
O469 [凝聚态物理学];
学科分类号
070205 ;
摘要
The sensitivity of continuous-wave NMR can be increased with a high-Q resonance tank circuit by using superconducting wire and a GaAs FET preamplifier operating at liquid helium temperatures. We discuss impedance matching to obtain an optimal signal-to-noise ratio for input circuits operating at almost-equal-to 500 kHz, with Q almost-equal-to 10(3)-10(4) at an impedance level of 1 MOMEGA and a noise voltage of 1.3 nV/square-rootHz. High rf compensation of 1:10(5) is achieved with orthogonally oriented transmitter coils and an astatically wound pick-up coil. The performance is evaluated on the basis of resolution that we achieved in the measurement of the NMR line in normal liquid He-3; this is better than 1 part in 10(3) of the peak amplitude, while in the integrated intensity the standard deviation from the mean is better than 0.3%.
引用
收藏
页码:159 / 160
页数:2
相关论文
共 5 条
[1]   NOISE MEASUREMENTS ON SILICON J-FETS AT LOW-TEMPERATURE USING A VERY HIGH Q-SUPERCONDUCTING RESONATOR [J].
AYELA, F ;
BRET, JL ;
CHAUSSY, J .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (11) :2816-2821
[2]   LOW-NOISE PREAMPLIFIERS FOR 1K OPERATION USING GALLIUM-ARSENIDE MESFETS OF VERY LOW 1/F NOISE [J].
CAMIN, DV ;
PESSINA, G ;
PREVITALI, E ;
RANUCCI, G .
CRYOGENICS, 1989, 29 (08) :857-862
[3]   OPTIMUM GEOMETRY OF SADDLE SHAPED COILS FOR GENERATING A UNIFORM MAGNETIC FIELD [J].
GINSBERG, DM ;
MELCHNER, MJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1970, 41 (01) :122-&
[4]  
LEPAISANT J, 1984, REV SCI INSTRUM, V55, P521, DOI 10.1063/1.1137784
[5]  
PARTS U, 1994, PHYSICA B, V194