摘要
The VITAL system principle is based on homogeneous fluorescence technology, During an Ii-month period, a total of 19,706 blood cultures from adult patients hospitalized in various establishments of the Montpellier Teaching Hospital were collected in VITAL bottles, of which 1,939 were declared positive, Only 204 bottles (1.04%) were false positives, The 1,735 true-positive bottles were collected from 130 patients, The final visual control permitted the detection of 10 falsely negative bottles (0.05%), of which 5 contained clinically significant microorganisms from four patients, The kinetics of detection for all microorganisms showed that 66.6% were detected within 24 h, 83.1% within 48 h, 95.5% within 120 h, and 100% within 150 h, No clinical episode would have been missed had a 5-day protocol been used instead of a 7-day protocol, Among the positive bottles, 65.7% were detected by the SLOPE algorithm, 20.1% by the DELTA algorithm, and 14.2% by the THRESHOLD algorithm, This retrospective study of our results shows that a 5-day protocol is sufficient for the detection of septic episodes using the VITAL system.
引用
收藏
页码:2098 / 2101
页数:4
相关论文
共 24 条
[1]   IMPORTANCE OF BLOOD-VOLUME CULTURED IN THE DETECTION OF BACTEREMIA [J].
ARPI, M ;
BENTZON, MW ;
JENSEN, J ;
FREDERIKSEN, W .
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 1989, 8 (09) :838-842
[2]   CONTAMINANT BLOOD CULTURES AND RESOURCE UTILIZATION - THE TRUE CONSEQUENCES OF FALSE-POSITIVE RESULTS [J].
BATES, DW ;
GOLDMAN, L ;
LEE, TH .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1991, 265 (03) :365-369
[3]  
BISMUTH M, 1994, LETT INFECTIOL, V13, P423
[4]  
BOUVET A, 1988, REV FR LAB, V179, P79
[5]   CLINICAL IMPLICATIONS OF POSITIVE BLOOD CULTURES [J].
BRYAN, CS .
CLINICAL MICROBIOLOGY REVIEWS, 1989, 2 (04) :329-353
[6]   BIOARGOS - A FULLY AUTOMATED BLOOD CULTURE SYSTEM [J].
COURCOL, RJ ;
DUHAMEL, M ;
DECOSTER, A ;
LEMAIRE, VM ;
RASTORGOUEFF, ML ;
OCHIN, D ;
MARTIN, GR .
JOURNAL OF CLINICAL MICROBIOLOGY, 1992, 30 (08) :1995-1998
[7]  
CROIZE Y, 1994, B SOC FR MICROBIOL, V9, P186
[8]  
DORSHER CW, 1991, REV INFECT DIS, V13, P633
[9]  
GRILLOT R, 1989, 34TH JOURN IT BIOL 2
[10]   TIME TO DETECTION OF POSITIVE BACT ALERT BLOOD CULTURES AND LACK OF NEED FOR ROUTINE SUBCULTURE OF 5-DAY TO 7-DAY NEGATIVE CULTURES [J].
HARDY, DJ ;
HULBERT, BB ;
MIGNEAULT, PC .
JOURNAL OF CLINICAL MICROBIOLOGY, 1992, 30 (10) :2743-2745