AVOIDING PERIPHERAL-NERVE STIMULATION - GRADIENT WAVE-FORM CRITERIA FOR OPTIMUM RESOLUTION IN ECHO-PLANAR IMAGING

被引:27
作者
HARVEY, PR [1 ]
MANSFIELD, P [1 ]
机构
[1] UNIV NOTTINGHAM,CTR MAGNET RESONANCE,DEPT PHYS,NOTTINGHAM NG7 2RD,ENGLAND
关键词
NEUROSTIMULATION; EPI;
D O I
10.1002/mrm.1910320213
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Adaptation of Mansfield and Harvey's theory of magnetically induced neural stimulation indicates that, for trapezoidal gradient waveforms and at the relatively low modulation frequencies commonly used in echo-planar imaging, an increase in rise time even to the point of creating a triangular waveform can prove advantageous over the use of simple sinusoidal modulation. For example, a higher image resolution is in principle possible before the neural stimulation limit is reached. For higher modulation frequencies and fast ramp trapezoidal waveforms, the peak gradient can be lower than that for a sinusoidal waveform by a factor approaching 2/pi for equal resolution. This means that by using a multimode resonant gradient drive, the amplifier power requirements are greatly relaxed. The signal sampling bandwidth required is also lower, because, due to the trapezoidal nonlinear sample pattern, sample points are spaced linearly during the waveform plateaus.
引用
收藏
页码:236 / 241
页数:6
相关论文
共 13 条
[1]  
BOURLAND JD, 1990, 9TH P SMRM ANN M, P1157
[2]  
BUDINGER TF, 1990, 9TH ANN M SOC MAGN R, P276
[3]   SENSORY STIMULATION BY TIME-VARYING MAGNETIC-FIELDS [J].
COHEN, MS ;
WEISSKOFF, RM ;
RZEDZIAN, RR ;
KANTOR, HL .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (02) :409-414
[4]   ANISOTROPY IN THE DIELECTRIC-PROPERTIES OF SKELETAL-MUSCLE [J].
EPSTEIN, BR ;
FOSTER, KR .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1983, 21 (01) :51-55
[5]   ELECTRICAL-CONDUCTIVITY OF SKELETAL-MUSCLE TISSUE - EXPERIMENTAL RESULTS FROM DIFFERENT MUSCLES INVIVO [J].
GIELEN, FLH ;
WALLINGADEJONGE, W ;
BOON, KL .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1984, 22 (06) :569-577
[6]   RESONANT TRAPEZOIDAL GRADIENT GENERATION FOR USE IN ECHO-PLANAR IMAGING [J].
HARVEY, PR ;
MANSFIELD, P .
MAGNETIC RESONANCE IMAGING, 1994, 12 (01) :93-100
[7]  
Hodgkin AL, 1967, CONDUCTION NERVOUS I
[8]   LIMITS TO NEURAL STIMULATION IN ECHO-PLANAR IMAGING [J].
MANSFIELD, P ;
HARVEY, PR .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (06) :746-758
[9]   MULTIMODE RESONANT GRADIENT COIL CIRCUIT FOR ULTRA HIGH-SPEED NMR IMAGING [J].
MANSFIELD, P ;
HARVEY, PR ;
COXON, RJ .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1991, 2 (11) :1051-1058
[10]  
PLONSEY R, 1988, BIOLECTRICITY QUANTI