TRANSMYOCARDIAL IMPEDANCE DURING SINGLE AND MULTIPLE INTERNAL VENTRICULAR DEFIBRILLATION SHOCKS

被引:2
作者
PETERS, W [1 ]
SOLINGEN, S [1 ]
KOBAYASHI, Y [1 ]
SCHARF, R [1 ]
MANDEL, WJ [1 ]
GANG, ES [1 ]
机构
[1] UNIV CALIF LOS ANGELES,CEDARS SINAI MED CTR,SCH MED,DEPT MED,DIV CARDIOL,LOS ANGELES,CA 90048
来源
AMERICAN JOURNAL OF PHYSIOLOGY | 1994年 / 267卷 / 02期
关键词
ELECTROPHYSIOLOGY; DIRECT CURRENT SHOCK; VOLTAGE; RESISTIVITY;
D O I
10.1152/ajpheart.1994.267.2.H684
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Little is known about the transmyocardial impedance during internal ventricular defibrillation. In a canine model, using high rate on-line digitization, random shock delivery, and titanium electrodes, we determined the relationship among voltage, current, and impedance, delivered energy, and defibrillation success within the individual and within successive defibrillation shocks. Impedance decreased with repeated defibrillation in 10 of 11 dogs. Impedance always increased during trapezoidal discharges, whereas voltage decreased. Impedance was lower with high energy-voltage shocks in all dogs. Visually, voltage and current waveform did not show a phase shift. There was no difference in the total energy delivered and the energy converted into heat by the resistive part of the impedance. With a formula valid only for resistive loads, the capacitance of the defibrillator was calculated to be within the measurement accuracy and tolerance of the factory-provided value of 132 mu F Polarization voltage was consistently observed. Thus the transmyocardial impedance during defibrillation is primarily resistive, nonlinear voltage dependent, and declines with successive shocks. Defibrillation success was not influenced by these phenomena.
引用
收藏
页码:H684 / H693
页数:10
相关论文
共 31 条
[1]   A NEW APPROACH TOWARDS DEFIBRILLATION ELECTRODES - HIGHLY CONDUCTIVE ISOTROPIC CARBON-FIBERS [J].
ALT, E ;
THERES, H ;
HEINZ, M ;
ALBRECHT, K ;
GEORG, H ;
BLOEMER, H .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11) :1923-1928
[2]  
BUDDE T, 1988, EUR HEART J, V9, P92
[3]   STRENGTH-DURATION CURVES OF FIXED PULSE WIDTH VARIABLE TILT TRUNCATED EXPONENTIAL WAVEFORMS FOR NONTHORACOMY INTERNAL DEFIBRILLATION IN DOGS [J].
CHAPMAN, PD ;
WETHERBEE, JN ;
VETTER, JW ;
TROUP, P ;
SOUZA, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (07) :1045-1050
[4]   THE POTENTIAL GRADIENT FIELD CREATED BY EPICARDIAL DEFIBRILLATION ELECTRODES IN DOGS [J].
CHEN, PS ;
WOLF, PD ;
CLAYDON, FJ ;
DIXON, EG ;
VIDAILLET, HJ ;
DANIELEY, ND ;
PILKINGTON, TC ;
IDEKER, RE .
CIRCULATION, 1986, 74 (03) :626-636
[5]  
Dahl C F, 1976, Med Instrum, V10, P151
[6]   THE RELATIONSHIP BETWEEN SUCCESSFUL DEFIBRILLATION AND DELIVERED ENERGY IN OPEN-CHEST DOGS - REAPPRAISAL OF THE DEFIBRILLATION THRESHOLD CONCEPT [J].
DAVY, JM ;
FAIN, ES ;
DORIAN, P ;
WINKLE, RA .
AMERICAN HEART JOURNAL, 1987, 113 (01) :77-84
[7]  
DEGARAVILLA L, 1981, 16TH AAMI ANN M WASH, P28
[8]   THE UP-AND-DOWN METHOD FOR SMALL SAMPLES [J].
DIXON, WJ .
JOURNAL OF THE AMERICAN STATISTICAL ASSOCIATION, 1965, 60 (312) :967-978
[9]   CARDIAC DAMAGE PRODUCED BY DIRECT-CURRENT COUNTER-SHOCK APPLIED TO THE HEART [J].
DOHERTY, PW ;
MCLAUGHLIN, PR ;
BILLINGHAM, M ;
KERNOFF, R ;
GORIS, ML ;
HARRISON, DC .
AMERICAN JOURNAL OF CARDIOLOGY, 1979, 43 (02) :225-232
[10]   DEFIBRILLATION CURRENT AND IMPEDANCE ARE DETERMINANTS OF DEFIBRILLATION ENERGY-REQUIREMENTS [J].
DORIAN, P ;
WANG, MJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1996-2001