RAY TRACING OF OPTICALLY MODIFIED FIBER TIPS .2. LASER SCALPELS

被引:25
作者
VERDAASDONK, RM
BORST, C
机构
[1] University Hospital Utrecht, Heart- Lung Institute, Utrecht, GA, NL-3508
来源
APPLIED OPTICS | 1991年 / 30卷 / 16期
关键词
D O I
10.1364/AO.30.002172
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The spatial irradiance distribution of tapered fibers and tapered rods used as scalpels in laser surgery has been calculated by ray tracing. The results were compared to measurements in air and in water. The beam profiles of laser scalpels were conically shaped at discrete angles related to the number of reflections within the scalpel. Light started to leak radially out of the scalpel before it reached the tip slightly suppressing the increase in fluence rate toward the tip. For effective tissue cutting, the scalpel tip may be shortened to optimize the irradiance increase in combination with radial energy leakage to obtain controlled hemostatic coagulation.
引用
收藏
页码:2172 / 2177
页数:6
相关论文
共 13 条
[11]   RAY TRACING OF OPTICALLY MODIFIED FIBER TIPS .1. SPHERICAL PROBES [J].
VERDAASDONK, RM ;
BORST, C .
APPLIED OPTICS, 1991, 30 (16) :2159-2171
[12]  
Welch A. J., 1985, HEAT TRANSFER MED BI, V2, P135, DOI DOI 10.1007/978-1-4684-8285-0_4
[13]   LASERS AND THE SURGEON [J].
WIEMAN, TJ .
AMERICAN JOURNAL OF SURGERY, 1986, 151 (04) :493-500