CONTINUOUS MONITORING OF MIXED VENOUS OXYGEN-SATURATION FOR DETECTING ALTERATIONS IN CARDIAC-OUTPUT AFTER DISCONTINUATION OF CARDIOPULMONARY BYPASS

被引:18
作者
INOMATA, S [1 ]
NISHIKAWA, T [1 ]
TAGUCHI, M [1 ]
机构
[1] UNIV TSUKUBA,INST CLIN MED,DEPT ANAESTHESIOL,TSUKUBA,IBARAKI 305,JAPAN
关键词
BLOOD; HEMOGLOBIN SATURATION; EQUIPMENT; FIBEROPTIC PULMONARY ARTERY CATHETER; OXIMETER; HEART; CARDIAC INDEX; CARDIAC OUTPUT; MONITORING; OXYGEN; CONSUMPTION; SATURATION;
D O I
10.1093/bja/72.1.11
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Although mixed venous oxygen saturation (S $($) over bar$$ v(o2)) is related to the reciprocal of cardiac output (CO) if both arterial oxygen content and oxygen consumption remain constant, simultaneous alterations in the three variables may occur immediately after discontinuation of cardiopulmonary bypass (CPB). To examine if continuous monitoring of S $($) over bar$$ v (O2) using a fibreoptic pulmonary artery catheter would be useful for detecting alterations in CO immediately after discontinuation of CPB, we have examined the relationships between changes in S $($) over bar$$ v(O2) cardiac index (Cl), oxygen consumption and haemoglobin concentration in 15 cardiac surgical patients. Changes in S $($) over bar$$ v(O2) correlated well with those in Cl for 60 min after CPB (r = 0.75, P = 0.0001). A closer correlation between these two variables was noted in the patients with baseline Cl of less than 2 litre min(-1) m(-2) (r = 0.87) than the patients with baseline Cl of more than 2 litre min(-1) m(-2) (r = 0.66). However, changes in S $($) over bar$$ v(O2) did not correlate with either oxygen consumption or haemoglobin concentration. The current results suggest that continuous monitoring of S $($) over bar$$ v(O2), with the fibreoptic pulmonary artery catheter may be useful for detecting changes in CO after discontinuation of CPB in patients with compromized cardiac function: (Br. J. Anaesth. 1994; 72: 11-16)
引用
收藏
页码:11 / 16
页数:6
相关论文
共 31 条
[11]   USE OF CONTINUOUS MONITORING OF MIXED VENOUS SATURATION IN THE CORONARY-CARE UNIT [J].
GORE, JM ;
SLOAN, K .
CHEST, 1984, 86 (05) :757-761
[12]   CONTINUOUS MONITORING OF MIXED VENOUS OXYGEN-SATURATION AS AN INDICATOR OF PHARMACOLOGIC INTERVENTION [J].
HASSAN, E ;
GREEN, JA ;
NARA, AR ;
JARVIS, RC ;
KASMER, RJ ;
POSPISIL, R .
CHEST, 1989, 95 (02) :406-409
[13]   THE VALUE OF MIXED VENOUS OXYGEN-SATURATION AS A THERAPEUTIC INDICATOR IN THE TREATMENT OF ADVANCED CONGESTIVE-HEART-FAILURE [J].
HASSAN, E ;
ROFFMAN, DS ;
APPLEFELD, MM .
AMERICAN HEART JOURNAL, 1987, 113 (03) :743-749
[14]   RELATION BETWEEN MIXED VENOUS OXYGEN-SATURATION AND CARDIAC INDEX - NONLINEARITY AND NORMALIZATION FOR OXYGEN-UPTAKE AND HEMOGLOBIN [J].
JAIN, A ;
SHROFF, SG ;
JANICKI, JS ;
REDDY, HK ;
WEBER, KT .
CHEST, 1991, 99 (06) :1403-1409
[15]  
JAMIESON WRE, 1982, CAN J SURG, V25, P538
[16]   PULMONARY-ARTERY CATHETER MIGRATION DURING CARDIAC-SURGERY [J].
JOHNSTON, WE ;
ROYSTER, RL ;
CHOPLIN, RH ;
HOWARD, G ;
MILLS, SA ;
TUCKER, WY .
ANESTHESIOLOGY, 1986, 64 (02) :258-262
[17]   INFLUENCE OF BALLOON INFLATION AND DEFLATION ON LOCATION OF PULMONARY-ARTERY CATHETER TIP [J].
JOHNSTON, WE ;
ROYSTER, RL ;
VINTENJOHANSEN, J ;
GRAVLEE, GP ;
HOWARD, G ;
MILLS, SA ;
TUCKER, WY .
ANESTHESIOLOGY, 1987, 67 (01) :110-115
[18]   MIXED VENOUS OXYGEN-SATURATION - ITS ROLE IN THE ASSESSMENT OF THE CRITICALLY ILL PATIENT [J].
KANDEL, G ;
ABERMAN, A .
ARCHIVES OF INTERNAL MEDICINE, 1983, 143 (07) :1400-1402
[19]   ONLINE MONITORING OF MIXED VENOUS OXYGEN-SATURATION AFTER CARDIOTHORACIC SURGERY [J].
KRAUSS, XH ;
VERDOUW, PD ;
HUGENHOLTZ, PG ;
NAUTA, J .
THORAX, 1975, 30 (06) :636-643
[20]   FIBEROPTIC REFLECTION OXIMETER [J].
LANDSMAN, MLJ ;
KNOP, N ;
KWANT, G ;
MOOK, GA ;
ZIJLSTRA, WG .
PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1978, 373 (03) :273-282