ROLE OF LOW-OSMOLALITY CONTRAST-MEDIA IN THROMBOEMBOLIC COMPLICATIONS - SCANNING ELECTRON-MICROSCOPY STUDY

被引:20
作者
CASALINI, E
机构
[1] Department of Radiology, Istituti Clinici di Perfezionamento, 20100 Milan
关键词
ANGIOGRAPHY; COMPLICATIONS; BLOOD; COAGULATION; CONTRAST MEDIA; EFFECTS; EMBOLISM;
D O I
10.1148/radiology.183.3.1584930
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Numerous in vitro studies have found that clot formation may occur when blood is mixed directly with nonionic low-osmolality contrast media during angiographic procedures because of activation of hemostasis in the catheter; ionic contrast media, on the other hand, inhibit clot formation. Thirty patients were injected with low-osmolality contrast media-15 with ioxaglate, an ionic dimer, and 15 with iopamidol, a nonionic monomer. The inner wall of the angiographic catheter was studied with electron microscopy after selective catheterization of supraaortic vessels. Clot formation of various extent was observed in nine (60%) of the patients administered iopamidol. No coagulation process was found in the catheters of the patients administered ioxaglate. Results indicate that nonionic, low-osmolality contrast media may play a role in the formation of thromboembolisms.
引用
收藏
页码:741 / 744
页数:4
相关论文
共 19 条
[1]   COMPLICATIONS OF CORONARY ARTERIOGRAPHY - FOLLOW-UP REPORT [J].
ADAMS, DF ;
ABRAMS, HL .
CARDIOVASCULAR RADIOLOGY, 1979, 2 (02) :89-96
[2]   EFFECT OF IONIC AND NONIONIC CONTRAST-MEDIA ON MORPHOLOGY OF HUMAN ERYTHROCYTES [J].
ASPELIN, P .
ACTA RADIOLOGICA-DIAGNOSIS, 1978, 19 (04) :675-687
[3]   SCANNING ELECTRON-MICROSCOPY OF SURFACE IRREGULARITIES AND THROMBOGENESIS OF POLYURETHANE AND POLYETHYLENE CORONARY CATHETERS [J].
BOURASSA, MG ;
CANTIN, M ;
SANDBORN, EB ;
PEDERSON, E .
CIRCULATION, 1976, 53 (06) :992-996
[4]  
CASALINI E, 1990, THROMBOEMBOLIC RISKS, P33
[5]  
DAWSON P, 1990, RADIOLOGY, V177, P280
[6]   THROMBOGENIC POTENTIAL OF NONIONIC CONTRAST-MEDIA [J].
FAREED, J ;
WALENGA, JM ;
SARAVIA, GE ;
MONCADA, RM .
RADIOLOGY, 1990, 174 (02) :321-325
[7]  
GRINES CL, 1989, CIRCULATION S, V80, P411
[8]   THROMBOEMBOLIC COMPLICATIONS IN CORONARY ANGIOGRAPHY ASSOCIATED WITH THE USE OF NONIONIC CONTRAST-MEDIUM [J].
GROLLMAN, JH ;
LIU, CK ;
ASTONE, RA ;
LURIE, MD .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1988, 14 (03) :159-164
[9]   SCANNING ELECTRON-MICROSCOPE STUDY ON HEMOSTATIC REACTION - MURAL THROMBUS AFTER REMOVAL OF ENDOTHELIUM, WITH SPECIAL REFERENCES TO PLATELET BEHAVIOR, SITE OF FIBRIN FORMATION AND MICROHEMOLYSIS [J].
HATTORI, A ;
WATANABE, T ;
IZUMI, T .
ARCHIVUM HISTOLOGICUM JAPONICUM, 1978, 41 (03) :205-227
[10]  
HWANG MH, 1988, J AM COLL CARDIOL, V11, P55