ELECTRO-THERMAL ATOMIZATION FROM METALLIC SURFACES .1. DESIGN AND PERFORMANCE OF A TUNGSTEN-TUBE ATOMIZER

被引:45
作者
SYCHRA, V [1 ]
KOLIHOVA, D [1 ]
VYSKOCILOVA, O [1 ]
HLAVAC, R [1 ]
PUSCHEL, P [1 ]
机构
[1] RES INST BROWN COAL, CS-43437 MOST, CZECHOSLOVAKIA
关键词
D O I
10.1016/S0003-2670(01)83757-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A tungsten-tube atomizer has been developed and tested for a number of elements. The atomizer can be accommodated in the support of a commercial Varian carbon rod atomizer (CRA-63 or CRA-90) and operated as an alternative to the graphite tubes and cups. Characteristic concentrations, detection limits, optimum atomization temperatures, reproducibility and profiles of calibration curves obtained for Ag, Al, Ba, Cd, Co, Cu, Mg, Mn, Ni, Pb and V are similar to those obtained with the graphite-tube atomizer. The main advantage of this atomizer is its simple construction, low cost, long lifetime and excellent analytical performance, particularly for elements forming refractory carbides. © 1979.
引用
收藏
页码:263 / 270
页数:8
相关论文
共 45 条
[21]  
LAGESSON HV, 1974, MIKROCHIM ACTA, P527
[22]   APPLICATION OF ELECTRODEPOSITION TECHNIQUES TO FLAMELESS ATOMIC-ABSORPTION SPECTROMETRY .1. DETERMINATION OF CADMIUM WITH A TUNGSTEN FILAMENT [J].
LUND, W ;
LARSEN, BV .
ANALYTICA CHIMICA ACTA, 1974, 70 (02) :299-310
[23]   APPLICATION OF ELECTRODEPOSITION TECHNIQUES TO FLAMELESS ATOMIC-ABSORPTION SPECTROMETRY .2. DETERMINATION OF CADMIUM IN SEA-WATER [J].
LUND, W ;
LARSEN, BV .
ANALYTICA CHIMICA ACTA, 1974, 72 (01) :57-62
[24]   APPLICATION OF ELECTRODEPOSITION TECHNIQUES TO FLAMELESS ATOMIC-ABSORPTION SPECTROMETRY .3. DETERMINATION OF CADMIUM IN URINE [J].
LUND, W ;
LARSEN, BV ;
GUNDERSEN, N .
ANALYTICA CHIMICA ACTA, 1976, 81 (02) :319-324
[25]  
MANSFIELD JM, 1974, TALANTA, V21, P787
[26]   INTERFERENCES OF ACIDS IN ATOMIC-ABSORPTION SPECTROMETRY WITH A TANTALUM FILAMENT ATOMIZER [J].
MARUTA, T ;
TAKEUCHI, T .
ANALYTICA CHIMICA ACTA, 1972, 62 (02) :253-259
[27]  
MARUTA T, 1973, BUNSEKI KAGAKU, V22, P602
[28]   FLAMELESS ATOMIC-ABSORPTION DETERMINATION OF COBALT,NICKEL, AND COPPER - COMPARISON OF TANTALUM AND MOLYBDENUM EVAPORATION SURFACES [J].
MCINTYRE, NS ;
COOK, MG ;
BOASE, DG .
ANALYTICAL CHEMISTRY, 1974, 46 (13) :1983-1987
[29]   FLAMELESS ATOMIC-ABSORPTION SPECTROMETRY EMPLOYING A WIRE LOOP ATOMIZER [J].
NEWTON, MP ;
DAVIS, DG .
ANALYTICAL CHEMISTRY, 1975, 47 (12) :2003-2009
[30]  
NEWTON MP, 1973, ANAL LETT, V6, P89